8. In the Event of Failure…
Trouble Shooting
Symptoms
Possible causes
Countermeasures
The POWER does
not turn ON.
ELB is turned OFF
Malfunction of the power supply
The wire ire short-circuited.
Malfunction of power switch
Turn the ELB ON
Check the power supply circuit
Replace the cord
Replace the power switch
The
touch
panel
operation screen will
not change even when
the unit is connected to
GAS410.
Imperfect connection of the connecting
cable
Connecting cable is defective
Connect correctly as per the
operation manual
Replace the cable
Blower does not
activate.
Incorrect
connecting
of
the
connector of blower
Lubricating oil level is low
Breaking of blower input cord
Blower switch failure
Blower or inverter failure
Blower circuit failure and wiring
failure
Connect correctly.
Exchange oil periodically
Replace the card.
Replace
the
touch
panel,
sequencer or thermo regulator.
Replace the blower or inverter
Maintain or replace the part
The freezer does
not operate.
The delay timer is in operation
Disconnection of the heater
Touch panel or sequencer is
defective
Freezer circuit and wiring are
defective
Wait for five minutes and check
whether the freezer operates or
not
Replace the heater
Replace the touch panel or the
sequencer
Repair the defective points or
replace the temperature controller
Spray pressure will
not increase.
Dial is set at 0
Touch panel or sequencer is
defective
Imperfect connection of the tube
Compressor is defective
Defective solenoid valve
Compressor circuit or wiring is
defective
Adjust the dial
Replace the touch panel or the
sequencer
Repair or replace the defective
points
Replace
Replace
Repair the defective point
The value of oxygen
concentration
monitor cannot go
beyond 18%
High altitude, low oxygen content
Flowmeter of oxygen sensor is
blocked
Pump failure
Filter degraded or blocking
Oxygen sensor failure
Oxygen sensor board failure
Use it at proper altitude or use
default oxygen concentration
Replace the flowmeter
Replace
Replace the filter
Replace
Replace
Oxygen
concentration
will
not decrease.
N2 supply source is defective
Flow meter for introducing N2 is
clogged
Imperfect connection of the tube
Defective solenoid valve
Touch panel or sequencer is
defective
Take care so that oxygen
concentration will not decrease
Replace the flow meter
Repair or replace the defective
points
Replace the solenoid valve
Replace the touch panel or the
sequencer