30
5. Handling Precaution
Caution during operation
(16)
If this unit is not operated, turn "OFF" the earth leakage breaker on the back of the unit.
(17)
The cyclone may charge easily with static electricity depending on the specific specimen used,
or operating environment or conditions. Implement countermeasures against static electricity
such as attaching included earth clips at three positions on the clamp at the connection of the
cyclone or attaching an antistatic brush to the body of the cyclone.
(18)
Depending on the sample to be processed, the static electricity may be occurred at cyclone.
Therefore, remove the static electricity with the appropriate method. It is efficient that the
wire is wounded to the glass portion for grounding, but it is more convenient to use the static
electricity remover by setting against the cyclone vertically.