21
4. Operating procedures
Operating method
KEY PANEL Description of indication lamps
You can confirm the operating statuses of switches one the KEY PANEL by checking whether a
specific lamp is on or off.
Each lamp will be turned on at the upper right corner on the KEY PANEL.
Lamp on: Indicates that the spray nozzle is attached.
Lamp blink: Indicates that a spray nozzle is not attached.
You can control temperature by setting a temperature on the outlet side temperature
controller while the lamp is on.
You can control temperature by setting a temperature on the inlet side temperature
controller while the lamp is on.
When the lamp is on, the blower is in operation.
When the lamp is on, the heater is in operation.
When the lamp is on, the liquid sending pump is operating in the normal direction.
When the lamp is on, the liquid sending pump is operating in the reverse direction.
When this lamp is on, it indicates that the freezer is operating.
When this lamp is flashing, the freezer is in the standby mode due to the delay
timer. The lamp shifts to lit status when the timer countdown finishes and the freezer
is activated.
When you want to abort processing of the sample, or when the nozzle is clogged
Abort sending liquid following operations of the ending process (P19
(
10
)~
P20
(
18
)
) when you
want to abort processing of the sample or the nozzle is clogged.
And if you want to process another sample, recover contents in the product collecting container, clean
it according to the maintenance method (P24." Cleaning After Using "), and then operate the unit
using another sample.
N
Inlet
B
H
R
outlet
FWD
REV