Operation
Yamatake Corporation
3-38
Model HGC303 - Heat Value Gas Chromatograph
(4) Total of raw concentration
Usually, the total of raw concentration is within 95-105 mol% during natural gas
analysis.
In these values are required to be changed, input a user defined value for both high and
low limit.
When these values exceed the limitations, the color of the “HGC Status” box turns to
red in user’s mode.
Default: High 105 mol%
Low 95 mol%
(5) Normalization method configuration
Normalization method can be selected.
Default is “Standard normalization”
(6) Multi-stream function setting
When you use HGC without multi-stream function, please select “Not use multi-
stream function” (default) and download to HGC to inactivate the multi stream
function.
When you use HGC with multi-stream function, please select “Use multi-stream
function” and download to HGC to activate the multi stream function.