Yamatake Corporation
Safety
Model HGC303 - Heat Value Gas Chromatograph
v
(2) Installation for FM Explosionproof / Flameproof Apparatus (in
accordance with NEC)
CAUTION
•
Install the apparatus only in hazardous (classified) locations for which the
apparatus has been approved.
•
Seal each conduit entering the apparatus enclosure within 18 in.(457 mm)
from the enclosure.
•
Do not open the apparatus enclosure when an explosive atmosphere is
present.
Figure S-1 An example of conduit seal (with stopping plug)
Figure S-2 An example of conduit seals (without stopping plug)
Summary of Contents for HGC303
Page 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Page 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Page 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Page 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Page 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Page 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...