Operation
Yamatake Corporation
3-54
Model HGC303 - Heat Value Gas Chromatograph
3-5-5 : Configuration mode
Numerous configurations of the model HGC303 and the HGM can be done in this mode.
Click on [Configuration mode] from the main menu. The following screen appears
after entering the password.
Figure 3-21 Configuration mode display (GPA)
Table 3-25 Description of configuration mode display (GPA)
Display
Description
Outputs configuration
Low cut off: Threshold of a peak or a noise
When the detected value < low cut off value, the output is
transmitted as 0 mol%.
%DEV RF Limit: See 3-4-2 Calibration procedure on page
3-20 for details.
Each RF %dev limit can be set independently.
PV12 to 15, 19 and 20 can be configured to set which out-
puts will be transmitted to the host control system.
PV High / Low Alarm Configuration:
High / Low Alarm limit values can be entered for PV1-11,
PV12-15, PV19-20.
This setting is available for application of
HGC (version 3.1 or later),
HGM (version 4.70 or later)
and HDM (version 2.40 or later) combination.
Summary of Contents for HGC303
Page 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Page 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Page 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Page 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Page 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Page 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...