(2) PV12 - 20 configuration
PV12 to 15, 19 and 20 can be configured to set which outputs are transmitted to the
host control system.
Possible configurations of each PV are as follows.
Table 3-17 Possible configurations of PV12-20
PV12
SCV (real) (MJ/m3) [default]
or SCV (ideal) (MJ/m3)
or ICV (real) (MJ/m3)
or ICV (real) (MJ/m3)
Following setting is available for application of HGC (version 3.1 or
later), HGM (version 4.70 or later) and HDM (version 2.40 or later)
combination.or SCV(real)(kJ/m3)or SCV(ideal)(kJ/m3) or
ICV(real)(kJ/m3)or ICV(ideal)(kJ/m3)or SCV(real)(kWh/m3) or
SCV(ideal)(kWh/m3)or ICV(real)(kWh/m3)or ICV(ideal)(kWh/m3)
PV13
Density (real) (kg/m3) [default]
or Density (ideal) (kg/m3)
or Relative density (real)
or Relative density (ideal)
PV14
Wobbe index (real) (MJ/m3) [default]
or Wobbe index (ideal) (MJ/m3)
Following setting is available for application of HGC (version 3.1 or
later), HGM (version 4.70 or later) and HDM (version 2.40 or later)
combination.or Wobbe index (real)(kJ/m3)or Wobbe
index(ideal)(kJ/m3)or Wobbe index(real)(kWh/m3) or Wobbe
index(ideal)(kWh/m3)
PV15
Compression factor Zmix
PV16
Total of raw concentrations
PV17
Oven temperature
PV18
Carrier gas pressure
PV19
ICV (real) (MJ/m3) [default]
or ICV (ideal) (MJ/m3)
Following setting is available for application of HGC (version 3.1 or
later), HGM (version 4.70 or later) and HDM (version 2.40 or later)
combination.or ICV(real)(kJ/m3)or ICV(ideal)(kJ/m3)or
ICV(real)(kWh/m3)or ICV(ideal)(kWh/m3)
Summary of Contents for HGC303
Page 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Page 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Page 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Page 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Page 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Page 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...