Yamatake Corporation
Appendix
Model HGC303 - Heat Value Gas Chromatograph
A-9
Helium is not included in Total raw.
Normalization with (Total raw + Helium)
PV1(normalized) = PV1(raw)
×
100 / [Total raw + Helium]
…
PV11(normalized) = PV11(raw)
×
100 / [Total raw + Helium]
Helium (normalized) = Helium
×
100 / [Total raw + Helium]
ISO calculation of Calorific value, density, relative density, Wobbe index and
compression factor will be calculated using PV1-PV11(normalized) and Helium
(normalized).
4 Helium (constant) normalization method
Model HGC303 calculates the normalized concentration using total raw and Helium
concentration by following formula.
Helium
Fixed value is used for Helium.
Fixed value can be entered.
Total raw calculation
Total raw = sum of raw concentrations (PV1-PV11)
Helium is not included in Total raw.
Normalization with (Total raw + Helium)
PV1(normalized) = PV1(raw)
×
100 / [Total raw + Helium]
…
PV11(normalized) = PV11(raw)
×
100 / [Total raw + Helium]
Helium (normalized) = Helium
×
100 / [Total raw + Helium]
ISO calculation of Calorific value, density, relative density, Wobbe index and
compression factor will be calculated using PV1 - PV11 (normalized) and Helium
(normalized).
Summary of Contents for HGC303
Page 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Page 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Page 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Page 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Page 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Page 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...