SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
76
measurement depts. The method used to calculate the mean velocity in a vertical
depends on the number of point velocity measurements made within a vertical.
Methods implemented for determining mean station velocity in a vertical are based on
the following literature,
•
ISO 748 – 2007, Hydrometry - Measurement of liquid flow in open channels using
current meters or floats,
•
WMO-No. 1044, Volume I – Fieldwork, 2010.
The methods implemented in FlowTracker2 for determining the Mean Station Velocity
are defined in
Table 5-2 - Determining Mean Station Velocity
Method
Measurement Depth
Mean Velocity Equation
0.6
0.6 * depth
V
mean
= CF * V
0.6
0.2/0.8
0.8/0.2
0.2 * depth
0.8 * depth
V
mean
= CF * ((V
0.2
+ V
0.8
) / 2)
.2/.6/.8
.8/.6/.2
0.2 * depth
0.6 * depth
0.8 * depth
V
mean
= CF * ((V
0.2
+ 2*V
0.6
+ V
0.8
) / 4)
Ice 0.6
0.6 * effective depth
V
mean
= 0.92*V
0.6
(Correction Factor 0.92 can be changed by user)
Ice 0.5
0.5 * effective depth
V
mean
= 0.88*V
0.5
(Correction Factor 0.88 can be changed by user)
Ice 2/8
Ice 8/2
0.2 * effective depth
0.8 * effective depth
V
mean
= CF * ((V
0.2
+ V
0.8
) / 2)
Kreps 2-
Kreps 2+
0.0 (near surface)
0.62 * depth
V
mean
= CF * (0.31*V
0.0
+ 0.634*V
0.62
)
5 Point+
5 Point-
0.0 (near surface)
0.2 * depth
0.6 * depth
0.8 * depth
1.0 (near bottom)
V
mean
= CF * ((V
0.0
+ 3*V
0.2
+ 3*V
0.6
+ 2*V
0.8
+ V
1.0
) / 10)
6 Point+
6 Point-
0.0 (near surface)
0.2 * depth
0.4 * depth
0.6 * depth
0.8 * depth
1.0 (near bottom)
V
mean
= CF * ((V
0.0
+ 2*V
0.2
+ 2*V
0.4
+ 2*V
0.6
+ 2*V
0.8
+ V
1.0
) / 10)
Vertical
Velocity
Curve
Any number of points at
user-specified depths
V
mean
= CF * Integrated velocity
Measurements can be made in any order; they are sorted
by depth to calculate the integrated mean velocity. Repeat
measurements at the same depth are averaged prior to
calculating the integrated velocity.
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...