SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
280
Figure 14:37. Station Editing Dialog
A pop-up editing dialog will appear when the
Edit
button is pressed, shown in
. Within this dialog, the following items can be edited:
•
GPS Latitude
– GPS latitude if known from external source
•
GPS Longitude
– GPS longitude if known from external source
•
Station Type
– Change station type (LeftBank, RightBank, IslandEdge,
OpenWater, Ice). Please note: when changing station types, there are some
station type combinations that will result in an error and the software will be
unable to compute discharge. For example, inputting two Left Bank station types
for the section will result in an error. Please refer to your discharge measurement
method guide for specific rules regarding what station types are valid for your
measurement.
•
Velocity Method (enter Manual Velocity or None)
– the option exists to enter a
manual velocity if it is known at the station and not measured by the ADV. To
enter a manual velocity, click the
Change
button and select the “Enter Velocity”
option, and hit
Accept
. A new input box will appear in the main dialog to enter a
manual velocity value. To select
None
for the Velocity Method, select the “None”
option, and hit
Accept
.
•
Location
– Change the location at which the station was taken. Please note: if
location does not satisfy the
Maximum Spacing Change
criteria set in the
Settings Pane, a warning will appear.
•
Depth
– Change the water depth at the station. Please note: if the depth does not
satisfy the
Maximum Depth Change
criteria set in the Settings Pane, a warning
will appear.
•
Correction Factor
– Change the correction factor applied to the velocity value.
The correction factor is a fraction ranging from [-1:1].
•
Comment
– Change or add a comment associated with station
•
Remove Measurement
– Removing a station removes it from being used in the
discharge calculation. Please note: some rules apply to what types of stations
can be removed, and are specific to the velocity method used. For example, if the
original measurement were a 0.2/0.8 measurement, removing the 0.2
measurement will result in an automatic removal of the 0.8 measurement. If a
0.2/0.6/0.8 measurement is taken, removing the 0.6 measurement will result in
removing the 0.6 measurement only, as a 0.2/0.8 is still a valid measurement
type. However, for this 0.2/0.6/0.8 measurement, if the 0.2 measurement is
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...