SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
49
Tilt Angle Warning
3.4.3.5
The
Tilt Angle Warning
parameter is the maximum threshold placed on the tilt angle of
the wading rod.
a) An angle of 0
°
means the wading rod is vertical (as desired for an ideal mounting
position).
b) The default Tilt Angle Warning applied in FlowTracker2 is 5 degrees.
c)
To change the Tilt Angle Warning,
i).
Use the backspace key to clear current entry,
ii).
Type in the new value using the keypad.
d)
To
DISABLE
the Tilt Angle Warning,
i).
Use the backspace key to clear current entry until “Disabled” is displayed.
Discharge Settings
3.4.4
The
Discharge Settings
function contains parameters required for discharge
calculations, discharge uncertainty analysis and quality control process.
Discharge settings function is only applicable to the Discharge Mode. The General
Mode does not contain any discharge calculation or uncertainty analysis.
Discharge Settings
screen includes the following
parameters,
a) Discharge Equation,
b) Discharge Uncertainty,
c) Discharge Reference,
d) Max Station Q for Warning,
e) Max Depth Change for Warning,
f) Max Spacing Change for Warning.
g) 0.6 Method Depth
h)
To select a parameter,
i).
Use up or down scroll arrows keys to select a
parameter,
ii).
The parameter selected will be highlighted in
yellow.
i)
To navigate to Template Menu,
i).
Press the Left Soft Key.
Figure 3:14 - Discharge Settings
Properties
associated with the parameters in discharge settings are defined in
Table 3-10 - Discharge Settings - Properties
Parameter
Min
Max
Default
Units
Required
Max Station Q for Warning
0.01
100
10
%
Yes
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...