SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
83
Quality Control
Section 6.
Quality Control
(QC) is continuously performed on all variables collected during a
measurement in both Discharge and General Modes. The
are automatically reviewed at different stages of a measurement and if any values
exceed the expected criteria a warning is supplied.
6.1. Quality Control Parameters
Quality Control Parameters
used to analyze the variables collected during a
measurement are discussed in the following section. The quality control parameters with
its expected values are supplied in
. All quality control parameters can be
adjusted or disabled based on measurement site and or user requirements.
Table 6-1 - Quality Control Parameters
Parameter
Description
Expected
SNR
SNR
is the most important Quality Control parameter,
•
It measures the strength of the acoustic reflection
from particles in the water,
•
Without sufficient SNR, the FlowTracker2 cannot
measure velocity.
Ideally > 10 dB
Minimum ≥ 4 dB
Velocity
Standard
Error
σV
(velocity standard error) is a direct measure of the
accuracy of velocity data,
•
It includes the effects of turbulence in the river and
instrument uncertainty.
Typically <
0.01m/s (0.03
ft/s).
Higher in
turbulent
environment.
Boundary
Interference
Boundary
QC evaluates the measurement
environment for interference from underwater
obstacles,
•
FAIR
or
POOR
results may indicate significant
interference from an underwater obstacle.
•
BEST,
•
GOOD,
•
FAIR,
•
POOR.
Velocity
Spike
Filtering
Spikes
in FlowTracker2 velocity data are removed
using a spike filter,
•
Some spikes are common and no cause for
concern,
•
Too many spikes indicate a problem in the
measurement environment (e.g., interference from
underwater obstacles or highly aerated water).
Typically < 5% of
total samples.
Should be < 10%
of total samples.
Velocity
Angle
Angle
is the direction of the measured velocity
relative to the FlowTracker2 X-axis,
•
Used for discharge measurements only,
•
A good site should have small velocity angles,
Ideally < 20º
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...