O-rings
11.3.4
The FlowTracker2 handheld makes use of two O-ring seals and it is important that the
following guidelines are followed.
Housing
11.3.4.1
a) The handheld is designed to withstand temporary submersion, but is not intended
for underwater operation.
b) We do not recommend opening the handheld housing without specific instructions
from SonTek.
Battery Cap
11.3.4.2
a) The O-ring located on the battery cap prevents moisture entering the battery
compartment.
b) Inspection of the O-ring should be during
defects that may exist on the O-ring.
c) The batter cap, O-ring and battery compartment must be inspected for any
moisture, dirt or material that can damage the O-ring, battery or battery
connectors.
Condensation in FlowTracker2 Housing
11.3.5
Moisture in the air can potentially damage FlowTracker2 electronics if allowed to
condense inside the handheld housing.
a) The handheld is shipped with a desiccant pack inside to absorb moisture,
b) We do not recommend opening the handheld housing without specific instructions
from SonTek. If the housing has not been opened, the desiccant should not need
to be replaced.
Zinc Anodes for Corrosion Protection
11.3.6
When using the FlowTracker2 in salt water, additional precautions must be taken to
prevent corrosion.
a) A sacrificial zinc anode should be installed on the probe (attached to the metal
portion of the probe stem),
b) Anode condition should be inspected regularly, and the anode should be replaced
when necessary. Zinc anodes are available from SonTek,
c) To check anode condition, try to chip away part of the anode with a screwdriver. If
large portions of the anode easily fall away, the anode should be replaced,
d) The probe and cable should be thoroughly rinsed with fresh water after each use.
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...