SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
188
In the case where the handheld was submerged under water or where it is covered in
mud and need to be rinsed, the following guidelines should be followed,
a) Do not remove the probe cable or battery cap while the handheld is under water
or drenched.
b) Use a dry cloth and remove excess moisture on the outside of the handheld.
c) Remove the “yellow Jacket” from the handheld casing and apply a dry cloth to the
handheld casing.
d) Remove the probe cable and inspect if the connector pins are dry. This can
indicate if there was water ingress into the connector.
e) Remove the battery cap and battery cartridge. Inspect both the battery cap and
cartridge for moisture as this can indicate water ingress.
f) Remove batteries from battery cartridge and inspect for moisture.
g) Place handheld, battery cartridge, batteries and yellow jacket on a dry place (not
in direct sunlight) and allow to dry.
If water ingress was identified during inspection in the battery compartment or probe
cable connector contact SonTek. It is not recommended to use the instrument if there
is moisture in the battery compartment.
Transducers
11.3.2.2
Biological growth on the transducers does not affect velocity measurements, but can
decrease acoustic signal strength and potentially increase noise in velocity data in clear
water.
a) Periodic cleaning of the FlowTracker2 transducers may be needed to maintain
optimal performance in areas of high biological activity,
b) FlowTracker2 transducers are encapsulated in an epoxy that is impervious to
damage from barnacles or other types of growth,
c) To remove growth, simply clean with a cloth or stiff (non-metallic) brush. The
transducer epoxy is very durable and cannot be easily damaged except by direct
impact.
Cable Maintenance
11.3.3
The FlowTracker2 probe cable is often the most vulnerable part of the system.
a) The cable uses a durable polyurethane jacket that provides excellent long-term
wear and abrasion resistance,
b) Any cable is susceptible to damage and reasonable precautions should be taken,
c) Inspect the cable and all connectors on a regular basis,
d) This cable is highly noise-sensitive and should not be modified by the user.
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...