![Xylem SonTek FlowTracker2 User Manual Download Page 275](http://html1.mh-extra.com/html/xylem/sontek-flowtracker2/sontek-flowtracker2_user-manual_3888944275.webp)
SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
274
Category
Edit
Value
Description
will use measured
temperature in velocity
calculation. If value is
entered, this value will be
used.
Sound speed
Sound speed used in velocity
measurement. If left blank,
will use measured
temperature and entered
salinity in calculation of sound
speed used in velocity
calculation. If value is
entered, this value will be
used.
Mounting correction
Mounting correction (%)
applied to velocity value.
Quality
Control
Settings
SNR threshold
SNR threshold (dB). If mean
SNR at sample volume is
below this value, a warning
will be issued.
Standard error threshold
Standard error threshold (in
velocity units). If standard
error over all samples at each
station exceeds this
threshold, a warning will be
issued.
Spike threshold
Spike threshold (in counts). If
number of spikes at a station
exceeds this value, a warning
will be issued.
Maximum velocity angle
Maximum velocity angle (in
degrees). If mean angle of
flow with respect to ADV X-
axis exceeds this value at a
station, a warning will be
issued.
Maximum tilt angle
Maximum tilt angle (in
degrees). If mean tilt at a
station exceeds this value, a
warning will be issued.
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...