133
0.8
Selecting the same fractional depth that was measured at the previous station improves
the overall efficiency of the measurement. The example that is shown above shows a
measurement where two-point method is performed at each station, keeping the order of
measurements synced between stations improves the workflow.
Step 3
Setting Fractional Depth
screen allow the setting
and placement of the wading rod before velocity
measurement is started,
a) Wading rod setting,
b) Beam Check,
c) SNR,
d) Tilt,
e) Velocity angle.
f)
To start a velocity measurement,
i).
Press the Right Soft Key.
g)
To
REVERT
to measurement screen,
i).
Press the Left Soft Key.
Figure 8:22 - Setting
Fractional Depth
The key features of the
Setting Fractional Depth
are explained in
Top Setting
wading rod
Automated Beam
Check
Wading rod
setting
Figure 8:23 - Features of Setting Fractional Depth
SNR
Tilt Bubble
Velocity Angle
Sampling
Volume
The Sampling Volume of the instrument is defined by the white vertical line displayed
on the Beam Profile plot in Figure 8:23. The location of the line is fixed and is based on
the geometry of the beams. This is an excellent guide to indicate if the boundary is too
close to the sampling volume.
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...