SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
173
Settings
9.3.4.1
The
Settings
menu option gives the user access to the initial measurement settings that
were configured during the configuration of the general template.
Settings screen consists of the following functions for
configuring measurement settings,
a) File Properties,
b) Data Collection Settings,
c) Quality Control Settings.
d)
To select a function,
i).
Use the up or down arrow key to scroll through
the functions,
e)
To
CLOSE
the settings menu,
i).
Press the Left Soft Key.
Figure 9:27 - Settings Menu
The measurement settings functions supplied in the Settings option in data collection
menu are defined under
Velocity Summary
9.3.4.2
Velocity Summary
from the data collection menu displays the mean velocity and other
variables measured from all completed stations. The velocity summary requires at least
one station with velocity measurements to populate the main variables in the table.
Velocity Summary
screen consists of tabular
summary of velocity and other variables measured up
to specific point,
b)
To close the velocity summary report,
i).
Press the Left Soft Key,
ii).
The software will navigate to the Data
Collection Window.
Figure 9:28 - Velocity Summary
Properties
associated with the variables and numerical calculations performed on the
raw data collected is defined in
Table 9-8 - Velocity Summary
Variable
Description
Units
Decimal
Vel. X Mean
Average station velocity, X velocity component
of all stations.
m/s or ft/s
4
Vel. X Min
Minimum station velocity, X velocity
component.
m/s or ft/s
4
Vel. X Max
Maximum station velocity, X velocity
m/s or ft/s
4
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...