131
Station Measurement (Wading Rod)
8.3.3
Station Measurement
is performed at each station that was created from the
subdivision of the measurements section into number of stations. The station
measurement consists of collecting a number of different parameters and variables
required for the Discharge Calculation Methods. Each station is unique with respect to
the selection of the location, width between adjacent stations, velocity method, averaging
interval used for velocity measurements and quality control parameters.
Station Parameters
8.3.3.1
The
Station Parameters
that need to be determined before a decision is made on what
velocity method will be used are the station location with respect to the starting point on
the tag line or measuring tape and the total water depth.
Figure 8:19 - Station Parameters
L
4
, is the distance from
the start location on the
tag line or measuring
tape.
D
4
, is the water depth
measured from the water
surface to the channel
bed.
The value of the start location on the tagline or measuring tape is not required to be
zero (0.000m). The station location can also be entered in negative direction, with the
start location at a larger value (e.g. 7.956m) and the end location at a smaller value
(e.g. 0.532m). This feature assist when a measurement is performed from the left bank
to the right bank and then back using the same tag line setup.
Start location of measurement at 0.532m
0.532
1.23
2.23
3.23
3.23
4.23
5.23
6.23
7.23
7.956
Start location of measurement at 7.956m
Point Velocity Measurement
8.3.3.2
The
Velocity Method
(see
Determining Mean Station Velocity
) selected for a station is
based on the water depth, hydraulic conditions and the flow distribution within the
measurement section.
•
The number of point velocity measurements required within a vertical is
dependent on the type of velocity method selected.
L
4
D
4
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...