Technical Data
148
POWER SUPPLY
Voltage
230V
~
(-15%/+10%) 50 ... 60 Hz
switchable
115V
~
(-15%/+10%) 50 ... 60 Hz
(Overvoltage category III)
Nominal power consumption
Model
Nominal power consumption
pH 171 (Standard)
13 VA
pH 171 R
15 VA
pH 171 RS
15 VA
pH 171 R/RS
20 VA
pH 171 2K
18 VA
pH 171 2K-R
20 VA
pH 171 2K-RS
20 VA
pH 171 2K-R/RS
23 VA
Behaviour after power failure:
Measurement is immediately continued as soon as mains voltage
is supplied again. Date and time have to be set!
EMC AND VDE STANDARDS
Protection:
IP 66, IEC 529, EN 60529, DIN VDE 0470
Protection class:
2 according to IEC 1010
Climate class:
Climate class 4,VDI / VDE 3540 Bl.2
Instrument security:
IEC 1010
EMC emission
EN 50081-1
EMC immunity
EN 50082-2, Namur recommendations: increased requirements
Test certificates:
GS, CE
Lightning protection
(data inputs and
High-voltage fuse limited to 90 V,
outputs)
Nominal leakage surge current 5 KA (8/20 µs)
fine protection max. 600 W
ELECTRICAL DATA
Channel A:
of low impedance, galvanically isolated, matching the output of
the SensoLyt
®
700-... armature, the adaptor ADA/pH 171
(preamplifier without temperature compensation and without
electrode check) and the connection box KI/pH 171 (preamplifier
with connection designed for temperature compensation and
temperature measurement).
Channel B:
of low impedance, galvanically isolated, matching the output of
the SensoLyt
®
700-... armature, the adaptor ADA/pH 171
(preamplifier without temperature compensation and without
electrode check) and the connection box KI/pH 171 (preamplifier
with connection designed for temperature compensation and
temperature measurement).
Models
-2K, -2K-R, 2K-RS,
with connection box KI 171
-2K-R/RS