Configuration
103
Selection of pH/mV channel B
This submenu is only valid for the P measuring mode. The submenu appears with CONF pH/mV
B and offers the same variants as for channel A.
In the D mode a unique configuration (pH/pH) is necessary because of the difference forming.
Channel B is automatically set to the selection of channel A.
Selection of calibration procedure
Chart:
Submenu
Setting possibilities
Conf KAL A
AUTOCAL TEC
*)
CONCAL
Conf KAL B
AUTOCAL TEC
*)
CONCAL
*) Ex-factory condition
In the CONF KAL menu the following calibration procedures can be selected:
AUTOCAL TEC
Automatic calibration with a selection of Technical buffer solutions
CONCAL
Conventional calibration with any buffers.
If pH was selected in the"CONF pH/mV A" menu the "CONF KAL A" menu will appear next.
Starting menu CONF KAL A.
With
select submenus.
With
go on to CONF T-C A.
Calls AUTOCAL TEC procedure.