Function as limit contact
If a relay is configured as a limit contact with UL / LL, free assignment to a parameter is possible.
The relay can be assigned to the following parameters:
Parameter
Parameter
Explanation
UL pH
A
LL pH
A
UL
= Upper limit value
UL mV
A
LL mV
A
LL
= Lower limit value
UL TEMP A
LL TEMP
A
pH
= pH value
UL pH
B
LL pH
B
mV
= Voltage measurement (Redox)
UL mV
B
LL mV
B
TEMP = Temperature
UL TEMP B
LL TEMP
B
A
= Channel A
UL pH
A-B
LL pH
A-B
B
= Channel B
UL mV
A-B
LL mV
A-B
A-B
= Difference A-B (only in D
UL TEMP A-B
LL TEMP
A-B
operation mode)
Parameters which have already been assigned are no longer offered in the menu during the
configuration of the next relay.
Selection make contact (close) / break contact (open)
In the submenu CONF REL O/C the function of a relay as make contact or break contact is laid
down.
Setting:
Submenu "CONF REL O/C“.
With
enter submenu
" CONF REL O/C ".
Relay 1 is closed in an active condition.
With
select between
CLOSE or
OPEN.
With
accept and go on to
setting of relay 2.
The relays 2 to 4 can be configured in the same way as make contacts or break contacts.
Exception: If a relay is assigned to the POWER FAIL function it always works as a break
contact. (With power failure, all relays are open!)
After the function for the last relay has been set the program goes back to the beginning of the
configuration menu; the display shows CONFIG.