Configuration
102
Selection of pH/mV channel A
The selection of the electrode lays down the possible functions and also the status indications of
the electrode supervision. The submenu has been assigned to both.
Indication
Submenu
CONF pH/mV A
pH electrode
pH
*)
Redox electrode
mV
*) Ex-factory condition
With pH the cullet supervision is permanently switched on,
with mV the cullet supervision is permanently switched off.
At one channel instruments only the variants "CONF pH/mV" appear. With
go to next
submenu.
Starting menu: pH/mV - channel A.
Submenu: pH - channel A.
With
select between pH and mV
(see next picture).
With
go on to CONF KAL A.
Submenu: mV - channel A.
With
select between pH and mV
(see previous picture).
With mV (Redox) NO CULLET
SUPERVISION.
With
go on to CONF T-P A.