- 5 -
Chapter 1 Overview
1.1 Scope of application
It is applicable to accurate, constant and continuous intravenous infusion in medical institutions.
1.2 Technical Parameters
IV Set Options /Size
: Open System for common widely-used brand of 10d/ml
、
15d/ml
、
20d/ml
、
60d/ml
(3.4mm
~
4.5mm), Also can be customized for dedicated IV set
Function Option
With Drug Lab.
Without Drug Lab.
Drug Lab.
Built-in 500 kinds of drugs, covering
the clinical use of the drug dose of 54
units on the library, and could be
connect with the PC for edit; export
& import.
NONE
Working Mode
Rate/VTBI
、
Time/VTBI
、
Rate/Time
、
Volume
、
Drop
、
Micro
、
Simple Rate
、
Programmable
、
DOSE Weight
、
Drug Lab
、
Auto Bolus
、
Manual Bolus.
IV Set
Open System for 10d/ml
、
15d/ml
、
20d/ml
、
60d/ml (3.4mm - 4.5mm)
IV Set management
could add / remove / edit more than 40 kinds tubes
Accuracy
Volume Mode
:
≤ ±5%
,
Drop Mode
:
≤ ±2% (after calibrate)
Calibrate
Automatic or manual calibration of infusion accuracy
Rate
Volume
:
0.1ml/h
~
1200ml/h
Purge
:
600ml/h
Drop
:
1d/min
~
400d/min
VTBI
1
~
9999 ml or OFF
Total Volume Σ
0
~
9999.9 ml
KVO
1ml/h
~
5ml/h Adjustable
,
Default
:
1ml/h
Programmable
Maximum 8 steps of different infuse rate / time / VTBI
BOLUS
Manual Bolus
:
long Press the Bolus key to infuse and release to stop
Auto Bolus
:
short press the Bolus key and set the VTBI & Rate
Upper Pressure
High / Middle / Low, 3 level, could off on the menu.
Down Pressure
200mmHg - 1200mmHg, 20 level, live pressure display, mmHg & kPa two units
Pre-Alarm Occlusion
When the pressure in the tube goes up, it has the function of early warning
Pre-Alarm Finish
1 min
~
30 min, The alarm can be set ahead of VTBI
Parameters Modify
changes the Rate, Time, Dose, VTBI & other important parameters during the infuse
working
Keypad Locker
Manual / Auto lock keyboard
History Record
Record 20000 infuse records, 10000 alarm records, can be downloaded to the computer