IV Set Installation
(
1
)
Under normal conditions, insert an ordinary PVC
IV set into the outlet of infusion bottle/bag.
(
2
)
Slightly squeeze Murphy’s dropper with fingers, to
make its 1/3 volume fill with liquid (see figure 9)
Notes
:
To ensure normal drip, never fill the dropper with water,
otherwise the infusion pump cannot detect drip drops and will cause
alarm; if the volume is less than 1/4, the bubble will easily go into
the infusion tube and cause bubble alarm.
Notes:
Please make sure that the air filter and inlet tube of infusion bottle/bag are smooth, otherwise it may
cause vacuum inside, then up occlusion happens, and prompt alarm.
Purge the air inside the infusion tube, make sure
no bubble inside, then close the adjustable clamp
of ordinary PVC set to prevent loss of liquid.
Open the infusion pump door, turn the locking
clamp (F) to the left to put the infusion tube.
Put the infusion set into guide slot A between
the liquid bottle/bag and the device.
Put the infusion tube into the guide slots B,C,D,E,
F, G and H vertically, so that the infusion tube can
be a straight line, as shown in figure 10.
Close the pump door tightly, then open the
infusion set adjustable clamp.
Notes
:
1. Make sure the infusion tube is put into the
bottom of ultrasonic sensor (guide slot C),
otherwise bubble alarm would happen.
2. Make sure no liquid drips in the Murphy’s dropper this moment (i.e. no leakage).