System parameters setup
System parameters in pause status
Press
.
Choose “system parameters”, press
.
Alarm setting
Sensitivity setting of up
occlusion detection
3 level options: high, median and low
When not easy to prompt occlusion alarm but it please
set to a higher level.
When easy to occlusion alarm but it prompts in a long
time or it doesn’t prompt, please set to a lower level.
Sensitivity setting of bottom
occlusion detection
20 level options: 200-1200mmHg
When not easy to occlusion alarm but it prompts,
please set to a higher level.
When easy to occlusion alarm but it prompts in a long
time or it doesn’t prompt, please set to a lower level.
Sensitivity setting of bubble
detection
When big bubble appears but alarm doesn’t prompt,
please choose the level of small number.
When bubble alarm prompt in high frequency, please
choose the level of larger number.
Pre-alarm setting of infusion
completed
Pre-alarm time setting of infusion completed: 5~30min
Can set to no need pre-alarm
Time setting of pause alarm
1min~24h settable
Fluid warmer
setting
Close fluid warmer
Choose not need heat
25~30 degrees centigrade
25~30 degrees centigrade
30~35 degrees centigrade
30~35 degrees centigrade
35~40 degrees centigrade
35~40 degrees centigrade
Bolus setting
Bolus rate setting
0.1~1200ml/h settable
Default 600ml/h
KVO setting
KVO rate setting
0.1~20ml/h settable
Key lock setting
Manual keypad lock setting
long press keypad lock to lock keypad after setting
Automatic keypad lock time
setting
Lock keypad automatically after 30s~1min settable
when infusion begins
Volume setup
3 level options
Choose suitable sound volume according to the environment