- 29 -
Chapter 6 General operation
Stop
Press
to stop infusion.
Power off
Press
to stop infusion.
Press
until the screen off.
Unplug the AC power cord.
Standby
Press “
” in stop status, until the screen turns dark.
Press any key in standby status, and the standby status would be removed.
(Piggyback)
Machine allows to modify the VTBI, Rate, Time, and to reset VI in process of infusion.
Choose the items need to be modified with up and down keys.
Input parameters, press
.
Notes
:
1
、
modify every two of VTBI, rate and time, another one would be calculated automatically.
2
、
when all these three parameters are changed, machine would calculate and infuse according to the
when last two modified parameters.
Purge
Press
in stop status, until the bubbles are purged totally.
Notes
:
Purge applies to expel the bubbles inside infusion tube, the infusion needle cannot be connected to
patient when execute purge operation.
Mute
Can suspend the alarm voice by using the Mute key when alarm prompts. The mute lasts 2 minutes, if new
alarm prompts would cancel mute automatically during the mute period, and prompt alarm voice.
Press
mute works, and the mute indicator light lightens.
Press
once again, cancel mute, alarm prompts again, indicator light goes out.
Clear alarm
When alarm prompts, should first exclude alarm reasons to know why alarm prompts.
Exclude alarm reasons and press
.
Notes
:
1
、
when AC power is not connected or AC power is interrupted, alarm prompts, can press
to
clear alarm, but the AC connection status icon on screen still prompts AC is not connected.
2
、
if AC power is connected, the alarm will be canceled automatically.
3
、
the alarm would prompt if there is no other operation when the pause time is up.
4
、
the pause time can be set to 2min~24h, the default is 2min.