- 35 -
Can close pressure pre-alarm if not needed
Abnormal down
pressure
High
Yes
Bottom pressure sensor is
interfered
Check for strong electromagnetic interference
around, re-start infusion pump after removal of
the source of interference
Bottom pressure sensor
fails
Pressure sensor is damaged, contact the
manufacturer for maintenance
Down Pressure
Sensor Failure
High
Yes
connection
of
bottom
pressure sensor looses
Contact the manufacturer for maintenance
IV set non
calibrated
High
Yes
IV set has not been
calibrated
Re-calibrate the IV set according to user’s manual
Abnormal rate
High
Yes
Tube type setting error
Not
re-set
the
rate
parameters after replacing
the tube
Set correctly the tube type
Set correctly the parameters
Drop Sensor
Failure
High
Yes
Connection of drop sensor
is off or looses
Reconnect the sensor
Connection is damaged
Contact the manufacturer for maintenance
Abnormal Drop
Sensor
High
Yes
Sensor is not properly
clipped
Clip it properly on Murphy’s dropper according
to User’s Manual
Leakage
High
Yes
Infusion tube is skew
Check the infusion line is installed correctly
Check whether the drip rate sensor is subjected to
strong interference/shaking
Check whether the door pressing plate is loose
Fluid Warmer
Failure
High
Yes
Fluid warmer connection
looses
Reconnect heater
Contact the manufacturer for maintenance
Battery
Failure “
”
High
Yes
Battery power is low and
cannot
be
charged
normally
Replace batteries
Battery off
High
Yes
Battery connection looses
Install the batteries correctly
Low battery
flashing)
low
No
Battery power is low
Connect AC supply to charge the battery
Motor error
High
Yes
Abnormal motor rotation
Contact the manufacturer for maintenance
Motor Stop
High
Yes
Stepper motor cannot run
normally
Infusion set is not properly installed, so that the
motor cannot drive the peristaltic pump normally.
Infusion pump fails, contact the manufacturer for
maintenance.
Pause Overtime
High
Yes
Infusion suspended over 2
min
Press "Clear" key to cancel the alarm and press
"Start" to continue the infusion
AC supply is
interrupted
low
No
AC
supply
is
not
connected correctly
Plug the AC supply correctly
Technical error
High
Yes
Device failure
Contact the manufacturer for maintenance
PS: if there is a single fault, the reason may be the infusion volume is less than 0.7m