- 32 -
Time setting
Set system time of machine
Can input year, month, date; hour, minute, second
TFT setting
Brightness adjustment of
LCD
Use up and down keys to adjust the brightness
System parameters check or setting in process of infusion
Press
.
Sensitivity setting of up
occlusion detection
Sensitivity setting of up
occlusion detection
3 level options: high, median and low
When not easy to prompt occlusion alarm but
it prompts, lease set to a higher level.
When easy to prompt occlusion alarm but it
prompts in a long time or it doesn’t prompt,
please set to a lower level.
Sensitivity setting of
bottom occlusion detection
Sensitivity setting of
bottom occlusion detection
20 level options:200-1200mmHg
When not easy to prompt occlusion alarm but
it prompts, please set to a higher level.
When easy to prompt occlusion alarm but it
prompts in a long time or it doesn’t prompt,
please set to a lower level.
Bubble setting
Sensitivity setting of bubble
detection
When big bubble appears but alarm doesn’t
prompt, please choose the level of small number.
When bubble alarm prompt in high
frequency, please choose the level of larger
number.
Pre-alarm setting of
infusion completed
Pre-alarm setting of
infusion completed
Pre-alarm time setting of infusion completed:
5~30min settable.
Can set to no need pre-alarm
Anti-bolus (Pressure
release)
The pressure of IV set
goes higher and the
pump will stop and the
motor will reverse to
reduce the pressure
Working with pressure sensor and control by the
pump software
Fluid warmer setting
Close fluid warmer
Choose not need heat
25~30 degrees centigrade
25~30 degrees centigrade
30~35 degrees centigrade
30~35 degrees centigrade
35~40 degrees centigrade
35~40 degrees centigrade
KVO setting
KVO rate setting
0.1~20ml/h settable
Battery life check
Check battery life
Check present battery run time