Chapter 10 Electrical specification
AC supply
AC power
AC supply
100 V - 240 V ~/50-60 Hz Class I
Maximum power
15VA
Fuse of protection
2 x 1A Fuse
Battery power supply
Battery
specification
9.6V 2000mA –nickel-metal hydride battery
Weight
About 150g
Battery runtime
More than 4 hours, in condition of infusion rate 25ml/h
Battery
charging time
Power on
:
12 h
Power off
:
10 h
Communication
Can communication wired or wireless communication
Certificate of Compliance
Comply with 93/42 / CE medical
directive
IPX4 Degree of Protection
(Water) - splash-proof
Degree of Protection
(Electrical) - Type CF
Type of Protection
(Electrical) - Class I
Medical electronic security equipment
Comply with EN / IEC 60601 - 1
and EN / IEC 60601-2-24
standards
EMC(electromagnetic
compatibility)
Comply with EN / IEC 60601 - 1
and EN / IEC 60601-2-24
standards
RoHS Certificate
Comply with EN/2002/95/EC
Revise EN/2011/65/EU directive