background image

- 3 -

Introduction

Thank you for choosing WIT-601A Infusion Pump. Please carefully read and understand this manual before
use. Regularly check and maintain the pump to avoid possible injury or risks.

Safety Tips

1. WIT-601A Infusion Pump is not explosion proof type or portable device.
2. WIT-601A Infusion Pump cannot be used in the condition with mixed gases of flammable anesthetic
gases, oxygen, or ammonia.
3. WIT-601A Infusion Pump can only be used by professional. Patients are not allowed to use.
4. Prevent patients from excessive or insufficient flow induced injury by the following methods: set
infusion parameters correctly, use the calibrated infusion device.
5. If there is any anomaly, suspend using should be taken as the preferred safety measure.
6. To avoid malfunction, keep the device away from strong electromagnetic source or noise when using,
such as NMR equipment, microwave generating devices, radiation equipment (X-ray, CT, etc).
7. To avoid malfunction, keep this device away from the high-frequency surgical instruments (such as
rack, cable, electrode feeder plates, etc.), the distance should be at least 25cm; at least 1m away from
the cell phone.
8. Do not use other voltages. Only use the specified voltages on nameplate of WIT-601A Infusion Pump,
or may cause damage or fire.
9. Do not throw batteries into fire or heat them, otherwise it will result in electrolyte leakage, fire, or
explosion.
10. Do not tear up the shell of batteries; otherwise it will cause an explosion or chemical burn hazard.
11. When plugging or unplugging the AC power, grip the plug firmly. Do not touch it with wet hands.
12. It is not recommended to share an outlet with other electrical equipments.
13. Never disassemble or modify the device arbitrarily.
14. Check the device daily. If not used for a long time, be sure that the functions are normal before
re-use.
15. If any anomaly or function loss, discontinue use and contact the supplier in a timely manner.
Otherwise the user shall take sole responsibility for any consequences.
16. Avoid vibration, bumps, direct sunlight or strong light.
17. Avoid direct contact with blowing hot, wet air from heating, electric stove or humidifier.
18. Avoid using the device in the following places: chemical storage place, dusty place,
vibration/shaking and damp places.
19. WIT-601A Series Infusion pump is class I in electrical safety classification. When in doubt in wiring
the protective earth conductor, it must be operated by the internal power supply.

20. WIT-601A Series syringe pump cannot be used in parallel with other brands of pumps.
21. The alarm sound pressure range of WIT-601A Series pump is 47-71 db, and the noise in the

operating environment cannot exceed 60 db.

22. Strictly in accordance with User’s Manual of infusion pump in use and maintenance.
23. Infusion pump using infrared drop sensor, does not apply to dark infusion tubing.

Summary of Contents for WIT-601A

Page 1: ...INFUSION PUMP WIT 601A User Manual Guangzhou WIT Medical Technology Co Ltd Version 3 0...

Page 2: ...pter 4 Operation 14 Function key description 14 Operation Flowchart 15 Volume mode operation 16 VTBI Rate Time 16 Rate Time 17 Simple Rate 17 Program mode 18 Dose weight mode 20 Drug library mode 21 C...

Page 3: ...Trouble shooting 34 Chapter 9 Routine check 37 Chapter 10 Electrical specification 38 Chapter 11 Storage and transport 39 User s Guide and Manufacturer s Statement of Anti electromagnetic Interference...

Page 4: ...em otherwise it will result in electrolyte leakage fire or explosion 10 Do not tear up the shell of batteries otherwise it will cause an explosion or chemical burn hazard 11 When plugging or unpluggin...

Page 5: ...different tubing type and different drug liquid it may cause large deviations in infusion accuracy 26 if alarms are presented the related medical personnel should be take the related solutions in time...

Page 6: ...cy Volume Mode 5 Drop Mode 2 after calibrate Calibrate Automatic or manual calibration of infusion accuracy Rate Volume 0 1ml h 1200ml h Purge 600ml h Drop 1d min 400d min VTBI 1 9999 ml or OFF Total...

Page 7: ...g Orange for low level alarm counting working Red for high level alarm pump stop Voice Tips Compliance YY 0709 2009 medical standard Alarm and key volume level adjustable with Mute function Monitor Hi...

Page 8: ...or components names description Infusion tube parts description Handle LCD Key Fluid warmer Drop sensor Infusion set guide slot Bubble sensor Bottom pressure sensor Infusion set guide slot Up pressure...

Page 9: ...supply status icon infusion status icon pressure status icon IV Set brand specification and fluid warmer temperature 2 Infusion parameter Machine status bar Battery status Infusion set info Flow rate...

Page 10: ...9 Infusion pump structure Figure 4...

Page 11: ...mbling Packing list Please check the follow contents when unpacking 1 Infusion pump body 1 set 2 Drop sensor 1 set 3 Power cord 1 set 4 User s manual 1 copy Pump Installation Horizontal installation o...

Page 12: ...11 Vertical installation on bracket Installation on cylinder...

Page 13: ...ugs into the rear sockets of WIT 601 series infusion pump according to the icon shown on pump back Notes if the drop sensor is not connected drop sensor failure will alarm 2 If not use the drop sensor...

Page 14: ...de then up occlusion happens and prompt alarm Purge the air inside the infusion tube make sure no bubble inside then close the adjustable clamp of ordinary PVC set to prevent loss of liquid Open the i...

Page 15: ...s CFM Set Used to confirm and save the items chose In the menu press it to enter next level menu Press it in process of infusion can reset infusion rate Menu Exit On the condition of pause press it to...

Page 16: ...r is normal etc after self check completed machine would enter infusion standby interface waiting for user to input infusion parameters Notes Not connect the AC power supply or the AC power is interru...

Page 17: ...interface after volume mode is chosen VTBI Rate Time Move the cursor to corresponding input boxes of rate VTBI and time with the up or down key Set rate Set VTBI Set time if need Press Notes set ever...

Page 18: ...I value would be displayed OFF Set rate Press Notes 1 Choose simple infusion mode should first move the cursor to VTBI input box with the up or down key long press for 1 seconds The VTBI value would b...

Page 19: ...f 1ml h Infusion mode selection Press in starting up status display infusion mode interface Choose program mode with up or down keys Press to enter program mode interface VTBI Rate Setting Programming...

Page 20: ...rogramming stage When no need more program infusion parameters press to enter the overview interface check whether the parameters of each programming stage are correct if some parameters of certain st...

Page 21: ...onding input boxes with the up or down keys Dose unit selection When dose unit flashes press to choose the dose needed with the up and down key Input dose value press or the down key Input weight Pres...

Page 22: ...Export Import Infusion mode selection Mode selection drug selection Press in starting up status display infusion mode interface Choose Drug Lib Mode with up or down key Press to enter the parameters...

Page 23: ...drug needed press At concentration input box press then the concentration will displayed N A which means the unknown drug concentration Input dose rate then press Input weight if needed Press Choose...

Page 24: ...process of infusion Set Bolus rate Set Bolus VTBI Press to begin Bolus Return to normal infusion after Bolus completed Manual Bolus Long press in process of infusion Machine would infuse according to...

Page 25: ...management Infusion set selection Brand selection Choose infusion set type Exit to main interface Press Choose IV Set menu press Choose Brand menu press Choose IV Set brand needed press Choose IV Set...

Page 26: ...measuring cup if choose electronic balance should put a vessel upon and reset the electronic balance before use 3 Close the adjustable clamp of infusion set waiting to be calibrated make sure no leaka...

Page 27: ...judge whether should calibrate again according to the actual value inputted Normally after 1 2 times calibration the accuracy would meet the requirement of deviation within 5 2 If alarm prompts in pro...

Page 28: ...brand of tube type actually used is not in the brand list of machine user can add infusion set brand by tube type add operation The operation as follows Choose menu Input brand Press confirm to finis...

Page 29: ...brand need be deleted Prompt delete or not Prompt deletion completed Choose Delete brand menu press Choose the IV Set brand need to be removed by using up and down keys press Choose according to the p...

Page 30: ...xpel the bubbles inside infusion tube the infusion needle cannot be connected to patient when execute purge operation Mute Can suspend the alarm voice by using the Mute key when alarm prompts The mute...

Page 31: ...menu of fluid warmer press Choose the menu of heating range press Press to return to infusion interface Pre alarm setting of infusion completion Press during infusion Choose the menu of infusion compl...

Page 32: ...appears but alarm doesn t prompt please choose the level of small number When bubble alarm prompt in high frequency please choose the level of larger number Pre alarm setting of infusion completed Pre...

Page 33: ...alarm but it prompts in a long time or it doesn t prompt please set to a lower level Bubble setting Sensitivity setting of bubble detection When big bubble appears but alarm doesn t prompt please choo...

Page 34: ...y records in machine Choose menu Choose item Input date Check results Check results Press Choose record menu press Choose the item need to be checked press Input the date need to be checked press Choo...

Page 35: ...d Check whether the infusion bottle bag is empty Check whether Murphy s dropper is misty or too tilted liquid flows down along the wall rather than dripping Up Occlusion High Yes Infusion inlet is not...

Page 36: ...heck the infusion line is installed correctly Check whether the drip rate sensor is subjected to strong interference shaking Check whether the door pressing plate is loose Fluid Warmer Failure High Ye...

Page 37: ...36...

Page 38: ...infusion set then boot Open pump door door open alarms and no liquid drops Close pump door and press clear key door open alarm disappears and no liquid drops Before boot Drop sensor Correctly install...

Page 39: ...ower on 12 h Power off 10 h Communication Can communication wired or wireless communication Certificate of Compliance Comply with 93 42 CE medical directive IPX4 Degree of Protection Water splash proo...

Page 40: ...Direct sunlight or strong light 2 Direct contact with blowing hot wet air from heating electric stove or humidifier 3 Chemical storage place or noxious gas place 4 Place of Water seepage splash dusty...

Page 41: ...Manufacturer s Statement of Anti electromagnetic Interference apply to all the equipments and systems WIT 601A Infusion Pump is intended to be used in the following specified electromagnetic environme...

Page 42: ...pment and systems WIT 601A Infusion Pump is intended to be used in the following specified electromagnetic environments Buyers or users should ensure that the pump is used in such environments Immunit...

Page 43: ...ion distance m based on communication equipment frequency 150KHz to 80MHz 80MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 As to the...

Page 44: ...inimum bias in infusion pump The testing programs for these results have been included in EN60601 2 24 IV set Drag Heart 20 drops ml standard PVC set The following graphs and curves show the overall c...

Page 45: ...compliance The equipments connected with WIT 601 should comply with their respective IEC or ISO standards i e information processing equipment IEC60950 and the whole equipment should comply with the...

Page 46: ...Address Unit 1103 Area C3 Innovation Building No 182 Science Road Science City High tech Industrial Development Zone Guangzhou China Tel 86 20 82118816 82118829 Fax 86 20 82118966 Web http www witmed...

Reviews: