Chapter 4 Operation
Function key description
Function keys concentrate upon the front housing, operated easily, the functions are as follows:
Function key
Description
On/Off
Used for boot and shutdown, long press for 1 second to boot, and long press 1 second to shut
down on the condition of no infusion.
Mute
Used when alarm happens, press it, the top left indicator of Mute lights up, and then the mute
status will last 2 minutes. Meanwhile it has mute key function, lasts 2 minutes.
Purge/BOLUS
On the condition of pause, long press to purge air inside tube; in process of infusion, press it
to enter BOLUS setting interface, then can start BOLUS infusion after setting parameters.
CFM/Set
Used to confirm and save the items chose. In the menu, press it to enter next level menu.
Press it in process of infusion, can reset infusion rate.
Menu/Exit
On the condition of pause, press it to enter the menu for parameter setting; and press it to
return to the previous menu; on the condition of standby, press it to exit the standby.
Start/Stop
Press it to start infusion on the condition of pause; and press it to stop infusion in process of
infusion.
Clear
On the condition of alarm, press it to cancel the sound-light alarm; on the condition of
inputting parameters, press it to set parameters to 0.
Lock/Dot
In process of infusion, long press for 1 second to lock or unlock. On the condition of pause,
long press to enter standby status.
Up
Used to move the cursor upwards.
Down
Used to move the cursor downwards.
AC
indicator
Used to indicate AC power connection.