Chapter 9 Routine check
In routine use, please perform routine maintenance and check according to the following
table:
Checkpoints
Check procedure
Check
stage
Cracks,
broken,
deformed
Check whether the housing of WIT-601 series infusion pump is intact, whether
cracks, broken and deformed exist.
Before use
AC status and AC
alarm
Boot and do not connect AC, the icon“ ” flashes and accompanies light-sound
alarm, by this moment connect AC, if the icon
appears, the AC supply is
normal, if not, the AC supply is abnormal.
boot
Finger-type
peristaltic pump
Correctly install infusion pump and clip infusion set, then boot.
Press “start” to start infusion. Check whether liquid drops or not, and check
whether the finger-type peristaltic pump turns abnormally.
In use
Bubble alarm
After booting, have not installed IV Set, but “bubble” alarms.
Install IV Set with liquid, close door, press “clear”, “bubble” alarm flash should
disappear.
In use
Pump door and
flow clip
Correctly install infusion pump and clip infusion set, then boot.
Open pump door, door-open alarms, and no liquid drops.
Close pump door and press “clear” key, door-open alarm disappears, and no
liquid drops.
Before boot
Drop sensor
Correctly install infusion pump and clip infusion set, then boot.
Start infusion and observe when every liquid drop drops, drop sensor would flash
one time.
In use
Abnormal drop
sensor
Take out drop sensor in process of infusion, stop infusion, prompting “drop
sensor falls off”.
Periodically
Occlusion alarm
Correctly install infusion pump and clip infusion set, then boot.
Set infusion rate 100ml/h, occlude manually the bottom infusion set after 5
minutes, and then stop infusion after about 1 minute, “down occlusion”
sound-light alarms.
Periodically
Infusion
accuracy
Correctly install infusion pump and clip infusion set, then boot, purge the bubbles
inside infusion set.
Set rate to 90ml/h, VTBI to 15ml, start infusion, measure the liquid
volume flowed out from machine with a measuring cup.
Check whether the liquid volume flowed out is within 15ml±5%.
Periodically
note
:
Carry out the above checks in specified using phase or check stage, if find error, please stop use and contact supplier as soon
as possible.