background image

6

3.2 Plumbing

The recommended plumbing for the instrument is shown in Figure 4. 

The instrument is designed to require very little head pressure to operate; 

around 6.9kPa (1 PSI). Flow requirements are 100ml/min. – 1.5 liter/min. 

(0.026-0.40Gal/min). The pressure regulator will allow for a maximum 

pressure of 1380kPa (200 PSI.). The maximum allowable fluid 

temperature is 50°C (122°F).
An optional flow regulator Cat# 19778 can limit flow to around 1 liter/

min. this is not recommended for applications involving sea water.

Figure 4: Recommended Plumbing for the Instrument

Intake Tubing 

Connection

4.75mm (

3

16

 inch)

I.D., 8mm (

5

16

 inch) 

O.D. tubing should 

be connected here 

to supply the sensor 

with a dependable 

sample flow

Drain Vent

Leave open for 

gravity-feed and 

open-drain systems. 

Plug with a #6-32 

sealscrew for  

pressurized systems. 

See manual for 

details

Backpressure 

Valve 

This valve allows 

adjustment of the 

amount of back 

pressure, which 

will help to control 

flow rate and 

eliminate small 

bubbles

Shutoff Clamp 

This clamp 

allows for 

shutoff of the 

intake flow 

during cuvette 

cleanings and 

replacements

Drain Tubing 

Connection

4.75mm (

3

16

 inch) 

I.D., 8mm (

5

16

 inch) 

O.D. tubing should 

be connected here 

to route the sensor 

drain tubing to a 

suitable site drain

Connector Assembly

Insert

Nut

Tubing  

(not supplied)

The instrument is equipped to be plumbed using 4.75 mm (

3

16

") ID, 

8 mm (

5

16

") OD flexible tubing. Opaque tubing should be used if the 

tubing will be exposed to sunlight, to prevent algae growth.

Model 28042 is equipped with a flow switch. For this instrument 

the user connections are equipped with ¼" push-to-connect 

connections. The suitable tubing is polymer or copper tubing 

with a ¼ inch OD. These connectors cannot be used with 

stainless steel tubing.

NOTICE

In figure 4, there are two operator controlled flow adjustment devices 

shown. The one on the input side is a shutoff clamp used during cuvette 

maintenance. The other device, located on the output side of the flow cell, 

is a backpressure valve. Backpressure may be required to prevent air from 

coming out of solution, which may be observed as tiny air bubbles.

3.2.1 Pressurized Systems

A standard feature of the AccUView LED is a pressure regulator. The 

regulator is rated for use up to 1380 kPa (200 PSI) and will lower the 

pressure down to a pressure more suitable for use in the AccUView 

LED. If inlet pressures can exceed 325 kPa (47 PSI) use appropriate 

tubing to connect to the AccUView LED. An optional flow regulator can 

be used to limit the flow to 1 liter per minute.

3.2.2 Drain Vent 

The AccUView LED has been fitted with a drain vent in the “OUT” 

bulkhead fitting. This fitting allows for atmospheric equalization, thus 

helping to alleviate bubble formation in the cuvette. Refer to Figure 4.
Upon initial flow minor leakage may occur through the drain vent. This 

will subside once normal flow is established.
For some high pressure systems, where the vent hole continuously 

leaks, a 6:32 seal screw is provided which should be inserted into the 

vent hole and tightened.

The drain water from this instrument should NEVER be reintroduced 

back into the incoming water stream. This water MUST be directed 

to a convenient drain.

NOTICE

Summary of Contents for HF scientific AccUView LED

Page 1: ...e Monitor IS HF AccUViewLED WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage o...

Page 2: ...d and passed EN 61326 1 2013 Manufacturer s Name HF scientific inc Manufacturer s Address 16260 Airport Park Drive Suite 140 Fort Myers Florida 33913 Importer s Name Importer s Address Type of Equipme...

Page 3: ...etting the Error Level 11 7 3 Configuring the RS 485 Port 11 7 3 1 Setting the Baud Rate 11 7 3 2 Setting the Address 11 7 4 Configuring the Alarms 11 7 4 1 Alarm 1 12 7 4 2 Alarm 2 12 7 5 Enabling th...

Page 4: ...indicates the items in the monitor shipment Item Quantity AccUView LED Monitor c w Field Terminal Box and Flow Through Assembly 1 Instruction Manual 1 Desiccant Pack 1 Quartz Cuvette with Ultrasonic...

Page 5: ...recommended mounting screws are M6 for the instrument enclosure and M4 3 16 for the field terminal box The AccUView LED is designed to have the field terminal box cradled under the sensor portion of...

Page 6: ...Model 28042 is equipped with a flow switch For this instrument the user connections are equipped with push to connect connections The suitable tubing is polymer or copper tubing with a inch OD These c...

Page 7: ...e While making connections refer to Figure 5 The AccUView LED is not supplied with a power cord 3 3 2 RS 485 The RS 485 Modbus or special protocol half duplex 2 wire digital interface operates with di...

Page 8: ...or the digital output depending on the options selected A flashing display may occur after warm up This is an indication that either the ambient or the water temperature or has changed rapidly During...

Page 9: ...e calibration 7 The upper display will count down the progress of the calibration step 8 The instrument will return to AUTO mode at the end the calibration 9 The shutoff clamp should now be released t...

Page 10: ...n of the AccUView LED will be resumed The AccUView LED will always reset to T and factory defaults after restoring NOTICE 7 0 Instrument Configuration CONFIG mode The instrument has been designed to p...

Page 11: ...rm 1 The alarm function HI LO OFF or Error 2 The alarm set point level at which the alarm activates 3 The delay time for the alarm the time that the set point must be exceeded prior to alarm activatio...

Page 12: ...tionalities is selected a prompt to set the delay times and the set point as with Alarm 1 will be displayed 7 5 Enabling the Security Access The instrument is equipped with a security access If this o...

Page 13: ...The factory 6 second update will extend the LED life to over 5 years The 60 second setting increases LED life to about 14 years but this will delay all updates Change the selected update period using...

Page 14: ...ust be used for the system to operate correctly The system works by sending an ultrasonic frequency through spring connections into a piezo transducer bonded to the bottom of a flow through cuvette Th...

Page 15: ...ed shutoff clamp unscrewing the old cuvette and replacing with a fresh clean one Recalibration will be required for accurate operation 9 2 Replacing or Installing the Desiccant Pouch The AccUView LED...

Page 16: ...the fault will be shown on the lower row of the display 10 2 System FAIL Message Normally this condition indicates that the instrument will require servicing Contact either the HF scientific Technica...

Page 17: ...ly 28474 AccUView LED Version with Flow Alarm Factory Installed Provides an alarm if sample flow is interrupted 28042 Tubing Kit Containing 1 shutoff clamp 1 backpressure valve 2 connecting tubings wi...

Page 18: ...3 500 87 33 3 438 30 17 1 188 147 66 5 813 Note 1 See the mounting instructions in the manual for mounting hardware sizes 2 Provide at least 200mm 8 inches of free space above the sensor for easy rem...

Page 19: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 20: ...ompany shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if t...

Reviews: