background image

12

7.4.1 Alarm 1

Alarm 1 Function:

 The 

ALM1

 is displayed and the display indicates 

the current function of alarm 1 

(HI, LO, OFF 

or

 Error)

. Use the 

 or 

 buttons to cycle through and select the desired function. Press the 

   

button to accept the selection.
If the alarm was turned 

OFF

 a prompt will appear to set up alarm 2  

(go to section 7.4.2). If, on the other hand, one of the other functionalities 

was selected a prompt will appear to set the delay times.

Alarm 1 Set Point:

 This prompt is used to select the set point for this 

alarm; this is indicated by “S/P” shown on the lower row of the display. 

Select the desired alarm level by using the 

 and 

buttons. Once the 

desired set point has been set, press the 

  button to accept it.

Alarm 1 Delay Times:

 

Delay On:

 The following display will appear to allow 

to select the number of seconds currently set for the “delay on” time.

The current selected number of seconds will be shown. Select the 

desired number of seconds for the “delay on” time for this alarm using 

the 

 and 

 buttons. Once the desired delay time has been set, press 

the 

  button to accept it.

Delay Off:

 Next, the following display will appear to select the number 

of seconds currently set for the “delay off” time.

The current selected number of seconds will be shown. Select the 

desired delay off time for this alarm using the 

 and 

 buttons. Once 

the desired delay time has been set, press the 

  button to accept it. 

After the settings for alarm 1 have been completed, prompts will allow 

for the set up of the information on alarm #2.

7.4.2 Alarm 2

Repeat the procedure listed in section 7.4.1 to set up the parameters 

for alarm 2. If one of the other functionalities is selected, a prompt 

to set the delay times and the set point, as with Alarm #1, will be 

displayed.

7.5 Enabling the Security Access

The instrument is equipped with a security access. If this option is turned 

on, the user is required to input the access code into the instrument to get 

to any mode other than 

AUTO

. The only code is 

333

. This code may not 

be changed. See section 4.3 for more information on this security feature. 

The security key icon will be visible and flashing on the display whenever 

the access option is selected using the 

 or 

 buttons. (

On 

or 

OFF

).

7.6 Extended Settings

The last few settings are grouped together to prevent them from being 

adjusted by accident. To gain access to the extended settings, select On 

using the 

 or 

 buttons and press the 

  button.

7.7 Units

Select the preferred measurement parameter in either ABS shown or %T. 

This selection will affect menus prior Extended Settings.

7.8 Speed of Response

The speed of response for both displayed and output values can be 

adjusted in this menu. The factory setting is 3 however, 10 response 

speeds are available. Select the desired speed of response using the 

 

and 

 buttons. Press the 

  button to accept it.

Please note that the response is simply an average of the previous 

readings combined with the current reading. If the setting is 1, only current 

readings are displayed. If the setting is 10, the last 9 readings are averaged 

with the current reading.

7.9 LCD Backlight Brightness

The LCD backlight brightness may need to be adjusted. This is of 

particular interest if multiple instruments are located in the same area and it 

is desired for the entire group to have the same appearance. Ten levels are 

available. The factory brightness is 8.
Change the brightness by pressing the 

 or 

 button. When the desired 

brightness has been selected, press the 

  button.

Summary of Contents for HF scientific AccUView LED

Page 1: ...e Monitor IS HF AccUViewLED WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage o...

Page 2: ...d and passed EN 61326 1 2013 Manufacturer s Name HF scientific inc Manufacturer s Address 16260 Airport Park Drive Suite 140 Fort Myers Florida 33913 Importer s Name Importer s Address Type of Equipme...

Page 3: ...etting the Error Level 11 7 3 Configuring the RS 485 Port 11 7 3 1 Setting the Baud Rate 11 7 3 2 Setting the Address 11 7 4 Configuring the Alarms 11 7 4 1 Alarm 1 12 7 4 2 Alarm 2 12 7 5 Enabling th...

Page 4: ...indicates the items in the monitor shipment Item Quantity AccUView LED Monitor c w Field Terminal Box and Flow Through Assembly 1 Instruction Manual 1 Desiccant Pack 1 Quartz Cuvette with Ultrasonic...

Page 5: ...recommended mounting screws are M6 for the instrument enclosure and M4 3 16 for the field terminal box The AccUView LED is designed to have the field terminal box cradled under the sensor portion of...

Page 6: ...Model 28042 is equipped with a flow switch For this instrument the user connections are equipped with push to connect connections The suitable tubing is polymer or copper tubing with a inch OD These c...

Page 7: ...e While making connections refer to Figure 5 The AccUView LED is not supplied with a power cord 3 3 2 RS 485 The RS 485 Modbus or special protocol half duplex 2 wire digital interface operates with di...

Page 8: ...or the digital output depending on the options selected A flashing display may occur after warm up This is an indication that either the ambient or the water temperature or has changed rapidly During...

Page 9: ...e calibration 7 The upper display will count down the progress of the calibration step 8 The instrument will return to AUTO mode at the end the calibration 9 The shutoff clamp should now be released t...

Page 10: ...n of the AccUView LED will be resumed The AccUView LED will always reset to T and factory defaults after restoring NOTICE 7 0 Instrument Configuration CONFIG mode The instrument has been designed to p...

Page 11: ...rm 1 The alarm function HI LO OFF or Error 2 The alarm set point level at which the alarm activates 3 The delay time for the alarm the time that the set point must be exceeded prior to alarm activatio...

Page 12: ...tionalities is selected a prompt to set the delay times and the set point as with Alarm 1 will be displayed 7 5 Enabling the Security Access The instrument is equipped with a security access If this o...

Page 13: ...The factory 6 second update will extend the LED life to over 5 years The 60 second setting increases LED life to about 14 years but this will delay all updates Change the selected update period using...

Page 14: ...ust be used for the system to operate correctly The system works by sending an ultrasonic frequency through spring connections into a piezo transducer bonded to the bottom of a flow through cuvette Th...

Page 15: ...ed shutoff clamp unscrewing the old cuvette and replacing with a fresh clean one Recalibration will be required for accurate operation 9 2 Replacing or Installing the Desiccant Pouch The AccUView LED...

Page 16: ...the fault will be shown on the lower row of the display 10 2 System FAIL Message Normally this condition indicates that the instrument will require servicing Contact either the HF scientific Technica...

Page 17: ...ly 28474 AccUView LED Version with Flow Alarm Factory Installed Provides an alarm if sample flow is interrupted 28042 Tubing Kit Containing 1 shutoff clamp 1 backpressure valve 2 connecting tubings wi...

Page 18: ...3 500 87 33 3 438 30 17 1 188 147 66 5 813 Note 1 See the mounting instructions in the manual for mounting hardware sizes 2 Provide at least 200mm 8 inches of free space above the sensor for easy rem...

Page 19: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 20: ...ompany shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if t...

Reviews: