background image

8

4.0 Operation

This process monitor allows for the measurement of the transmittance 

of process water on-line. Process water is usually reported in units 

of %T. Readings above 102 %T are indicated by a flashing display. 

Readings above 100 %T indicate that the current sample is better than 

the calibration water. Readings above 110 %T are not possible. All 

readings are path length corrected to 10 mm.
Alternately measurements can be taken in Absorbance. Negative 

Absorbance readings indicate the need for a calibration or that the 

water tested is better that what the instrument was calibrated to. These 

readings show ABS on the lower line of the display. All readings are 

path length corrected to 10 mm
During normal operation, the instrument will have the arrow beside 

AUTO

. The scale (% T or ABS) is displayed on the lower row of the 

display and the measured reading on the upper row of the display 

(see illustration below). The user needs to keep in mind 

that the instrument only updates at periods determined in 

the configuration or 

CONFIG

 menu. Refer to section 

7.14 

Lamp Duty Cycle

.

4.1 Warm-up

Upon power-up the AccUView LED will require a warm-up period of about 

30 minutes. For improved accuracy, allow the AccUView LED to compete 

warm-up time prior to calibrating. During the warm-up period, the display 

may flash indicating that it has detected a temperature change. It is normal 

for this to occur during the warm-up period.

4.2 Routine Measurement

The following steps describe how to measure the % T or ABS of a sample 

using this instrument:
1.  Apply power to the instrument and allow the unit an initial warm up 

of 30 minutes. 

2.  When a continuous process stream is flowing through the 

instrument, the instrument will display the measured % T or ABS 

level of the sample by displaying it on the LCD screen. In addition, 

the equivalent signal is provided on the analog (4-20 mA) output, or 

the digital output, depending on the options selected. 

A flashing display may occur after warm-up. This is an 

indication that either the ambient or the water temperature 

or has changed rapidly. During this time the readings may be 

slightly out of the specified accuracy. The AccUView LED will 

automatically compensate as soon as the rate of temperature 

change slows.

NOTICE

4.3 Security Access Feature

The instrument is equipped with a security access code feature that can 

be activated in the configuration mode. If the security feature is enabled, 

the screen shown in the illustration below will appear when the 

MODE/EXIT

 

button is pressed.

The security code (333) must be entered to gain access to 

CAL

 or 

CONFIG

 menus. Notice that the first number in the code is flashing. 

The flashing indicates that this is the number to be entered. Use the 

 or 

 arrows to select the first of the three numbers in the code and 

then press the 

  button to accept the first number of the code. Now 

enter the second number in the code. Proceed as with the first number 

followed by 

 . Then repeat the process for the third number in the 

access code, and finish with the 

  button.

If the valid access code has been selected, the instrument will be 

directed to the calibration mode. If the CONFIG mode is needed push 

the MODE/EXIT button one more time.

If the wrong access code is selected, the instrument will 

return to the AUTO mode. Refer to section 

7.6 Enabling the 

Security Access

 for more information.

4.4 Initial Operation

It is recommended to check that cuvette is tightly installed on the flow 

through unit prior to the introduction of water. These are pressure checked 

at the factory but due to shipping the cuvette could have loosened.
For the best accuracy an initial calibration is recommended. These are all 

factory calibrated but the best accuracy is achieved by calibrating with 

DI water with the flow through locked in place prior to calibration. Do not 

move the flow through after calibration, but simply start the fluid flow. Refer 

to the next section for complete calibration instructions.

Summary of Contents for HF scientific AccUView LED

Page 1: ...e Monitor IS HF AccUViewLED WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage o...

Page 2: ...d and passed EN 61326 1 2013 Manufacturer s Name HF scientific inc Manufacturer s Address 16260 Airport Park Drive Suite 140 Fort Myers Florida 33913 Importer s Name Importer s Address Type of Equipme...

Page 3: ...etting the Error Level 11 7 3 Configuring the RS 485 Port 11 7 3 1 Setting the Baud Rate 11 7 3 2 Setting the Address 11 7 4 Configuring the Alarms 11 7 4 1 Alarm 1 12 7 4 2 Alarm 2 12 7 5 Enabling th...

Page 4: ...indicates the items in the monitor shipment Item Quantity AccUView LED Monitor c w Field Terminal Box and Flow Through Assembly 1 Instruction Manual 1 Desiccant Pack 1 Quartz Cuvette with Ultrasonic...

Page 5: ...recommended mounting screws are M6 for the instrument enclosure and M4 3 16 for the field terminal box The AccUView LED is designed to have the field terminal box cradled under the sensor portion of...

Page 6: ...Model 28042 is equipped with a flow switch For this instrument the user connections are equipped with push to connect connections The suitable tubing is polymer or copper tubing with a inch OD These c...

Page 7: ...e While making connections refer to Figure 5 The AccUView LED is not supplied with a power cord 3 3 2 RS 485 The RS 485 Modbus or special protocol half duplex 2 wire digital interface operates with di...

Page 8: ...or the digital output depending on the options selected A flashing display may occur after warm up This is an indication that either the ambient or the water temperature or has changed rapidly During...

Page 9: ...e calibration 7 The upper display will count down the progress of the calibration step 8 The instrument will return to AUTO mode at the end the calibration 9 The shutoff clamp should now be released t...

Page 10: ...n of the AccUView LED will be resumed The AccUView LED will always reset to T and factory defaults after restoring NOTICE 7 0 Instrument Configuration CONFIG mode The instrument has been designed to p...

Page 11: ...rm 1 The alarm function HI LO OFF or Error 2 The alarm set point level at which the alarm activates 3 The delay time for the alarm the time that the set point must be exceeded prior to alarm activatio...

Page 12: ...tionalities is selected a prompt to set the delay times and the set point as with Alarm 1 will be displayed 7 5 Enabling the Security Access The instrument is equipped with a security access If this o...

Page 13: ...The factory 6 second update will extend the LED life to over 5 years The 60 second setting increases LED life to about 14 years but this will delay all updates Change the selected update period using...

Page 14: ...ust be used for the system to operate correctly The system works by sending an ultrasonic frequency through spring connections into a piezo transducer bonded to the bottom of a flow through cuvette Th...

Page 15: ...ed shutoff clamp unscrewing the old cuvette and replacing with a fresh clean one Recalibration will be required for accurate operation 9 2 Replacing or Installing the Desiccant Pouch The AccUView LED...

Page 16: ...the fault will be shown on the lower row of the display 10 2 System FAIL Message Normally this condition indicates that the instrument will require servicing Contact either the HF scientific Technica...

Page 17: ...ly 28474 AccUView LED Version with Flow Alarm Factory Installed Provides an alarm if sample flow is interrupted 28042 Tubing Kit Containing 1 shutoff clamp 1 backpressure valve 2 connecting tubings wi...

Page 18: ...3 500 87 33 3 438 30 17 1 188 147 66 5 813 Note 1 See the mounting instructions in the manual for mounting hardware sizes 2 Provide at least 200mm 8 inches of free space above the sensor for easy rem...

Page 19: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 20: ...ompany shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if t...

Reviews: