15

8.3.2 Simple Communication

The AccUView LED can provide basic communications over simple 

programs such as the Hilgraeve HyperTerminal that is included with 

most Microsoft Windows packages. The user could also use Visual Basic 

or other programs. The default communication parameters are 8 bits, 

no parity and 1 stop bit. These parameters may be changed in the 

extended 

CONFIG

 menu.

The master computer will send out:
• Byte #1 the attention character “:” in ASCII or 

3A

 Hex

• Byte #2 the address of the AccUView LED being queried
• Byte #3 and 4 CR LF or 0D 0A in hex
The AccUView LED will respond with:
• The same attention character “:” in ASCII or 

3A

 Hex

• The address of the AccUView LED
• The Reading
• The Unit (% T or ABS )
A sample communication would look like this:
(Master computer requesting a report from address #1)  

: 1 CRLF

(AccUView LED set to address #1 Response)  

:001 97.4 %T

8.3.3 Modbus Communication

Modbus communications is fully operational if the RS-485 option 

(Catalog #19851A) is installed. The Modbus protocol communication 

information is covered in a separate manual, Catalog #24570 and is 

available as a free download from our website at www.hfscientific.com.

8.4 AccUView LED with Flow Alarm (Model 28042)

This model of the AccUView 

LED has flow switch installed. 

There is an additional menu in the 

configuration which gives you the 

selection of ON or OFF.
A “Low Flow” occurs under 1 

Liter/minute and if the flow alarm 

is enabled will cause a screen 

indication. The flow alarm can 

act on both the relays and the 

4-20mA. Either relay, if set to Error 

will change to an alarm state and 

the 4-20 mA signal will change o 

0, 2 or 4 mA (selected by ERLV) 

unless it is set to OFF. A modbus 

register is also set. This alarm 

takes 2½ minutes of low flow to 

actuate or restore. Please note 

that the flow alarm is not sold as 

an option and cannot be added.

9.0 Routine Maintenance

9.1 Cleaning the Flow Through Cuvette

Measurement cuvettes used for the flow through should be clean and free 

of marks or scratches. Cleaning is accomplished by cleaning the interior and 

exterior with a detergent solution and then rinsing several times with distilled 

or de-ionized water. The cuvette can be replaced by first shutting off the flow 

using the provided shutoff clamp; unscrewing the old cuvette and replacing 

with a fresh clean one. Recalibration will be required for accurate operation.

9.2  Replacing or Installing the Desiccant 

Pouch

The AccUView LED continuously checks the condition of the desiccant. 

When the desiccant gets in such a condition that it may cause 

problems, the instrument will display 

DESC

 on the lower portion of the 

display to indicate the presence of humidity.

Proper use of the supplied desiccant is essential in maintaining 

the performance of the instrument. The desiccant has been 

designed to have a long life; however, replacement of the 

desiccant pouch will be required from time to time.

It is essential that the enclosure seal on the instrument base be 

maintained to ensure adequate desiccant life. Inspect the seal each 

time the desiccant pouch is replaced. Replace or reseat the seal if it is 

found to be defective.
The desiccant should be replaced when the instrument displays 

DESC

A new sealed desiccant pouch and indicator card are available from HF 

scientific part #21555R.To initially install or remove the old desiccant, simply 

unscrew the four corner thumbscrews and remove the electronics half of the 

instrument. Open the bag protecting the new desiccant pouch and replace 

(or install) immediately. To speed up the recognition, by the instrument, of the 

new desiccant it will be necessary to reset the instrument by disconnecting 

the sensor interconnect cable for 2 seconds and then reconnecting it.

Ensure this cuvette is tightened to the flow through before each 

operation. After a cuvette it is recommend that the water be run with 

the flow through not installed to allow for inspection of the seal.

NOTICE

Humidity Indicator

(Humonitor Card)

Replaceable 

Desiccant Pouch 

(Part 21555R)

Drain Seal

Center of drain 

opening in  

older models

Once the bag is opened, install the desiccant pouch immediately 

to prevent premature degradation of the desiccant.

NOTICE

9.3 Replacing the Source LED

The UV source lamp in the AccUView LED has a very long life, but is 

dependent on the setting of the Duty Cycle and operating conditions. 

See section 7.14 for more information on the LED life
When the UV source LED requires replacement it can easily performed 

but it does require opening the enclosure. Contact the factory before 

servicing. Alternately another service module can be purchased. Call 

the HF Customer Service Department for assistance.

Summary of Contents for HF scientific AccUView LED

Page 1: ...e Monitor IS HF AccUViewLED WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage o...

Page 2: ...d and passed EN 61326 1 2013 Manufacturer s Name HF scientific inc Manufacturer s Address 16260 Airport Park Drive Suite 140 Fort Myers Florida 33913 Importer s Name Importer s Address Type of Equipme...

Page 3: ...etting the Error Level 11 7 3 Configuring the RS 485 Port 11 7 3 1 Setting the Baud Rate 11 7 3 2 Setting the Address 11 7 4 Configuring the Alarms 11 7 4 1 Alarm 1 12 7 4 2 Alarm 2 12 7 5 Enabling th...

Page 4: ...indicates the items in the monitor shipment Item Quantity AccUView LED Monitor c w Field Terminal Box and Flow Through Assembly 1 Instruction Manual 1 Desiccant Pack 1 Quartz Cuvette with Ultrasonic...

Page 5: ...recommended mounting screws are M6 for the instrument enclosure and M4 3 16 for the field terminal box The AccUView LED is designed to have the field terminal box cradled under the sensor portion of...

Page 6: ...Model 28042 is equipped with a flow switch For this instrument the user connections are equipped with push to connect connections The suitable tubing is polymer or copper tubing with a inch OD These c...

Page 7: ...e While making connections refer to Figure 5 The AccUView LED is not supplied with a power cord 3 3 2 RS 485 The RS 485 Modbus or special protocol half duplex 2 wire digital interface operates with di...

Page 8: ...or the digital output depending on the options selected A flashing display may occur after warm up This is an indication that either the ambient or the water temperature or has changed rapidly During...

Page 9: ...e calibration 7 The upper display will count down the progress of the calibration step 8 The instrument will return to AUTO mode at the end the calibration 9 The shutoff clamp should now be released t...

Page 10: ...n of the AccUView LED will be resumed The AccUView LED will always reset to T and factory defaults after restoring NOTICE 7 0 Instrument Configuration CONFIG mode The instrument has been designed to p...

Page 11: ...rm 1 The alarm function HI LO OFF or Error 2 The alarm set point level at which the alarm activates 3 The delay time for the alarm the time that the set point must be exceeded prior to alarm activatio...

Page 12: ...tionalities is selected a prompt to set the delay times and the set point as with Alarm 1 will be displayed 7 5 Enabling the Security Access The instrument is equipped with a security access If this o...

Page 13: ...The factory 6 second update will extend the LED life to over 5 years The 60 second setting increases LED life to about 14 years but this will delay all updates Change the selected update period using...

Page 14: ...ust be used for the system to operate correctly The system works by sending an ultrasonic frequency through spring connections into a piezo transducer bonded to the bottom of a flow through cuvette Th...

Page 15: ...ed shutoff clamp unscrewing the old cuvette and replacing with a fresh clean one Recalibration will be required for accurate operation 9 2 Replacing or Installing the Desiccant Pouch The AccUView LED...

Page 16: ...the fault will be shown on the lower row of the display 10 2 System FAIL Message Normally this condition indicates that the instrument will require servicing Contact either the HF scientific Technica...

Page 17: ...ly 28474 AccUView LED Version with Flow Alarm Factory Installed Provides an alarm if sample flow is interrupted 28042 Tubing Kit Containing 1 shutoff clamp 1 backpressure valve 2 connecting tubings wi...

Page 18: ...3 500 87 33 3 438 30 17 1 188 147 66 5 813 Note 1 See the mounting instructions in the manual for mounting hardware sizes 2 Provide at least 200mm 8 inches of free space above the sensor for easy rem...

Page 19: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 20: ...ompany shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if t...

Reviews: