background image

5

2.4 Display

Figure 1 illustrates all the items that can appear on the display. The 

upper row of the display (1) is used for reporting the % T levels and to 

provide user guidance in the customer setting routine. The lower row 

of the display (2) is used to communicate error messages and provide 

user guidance. All error messages will be added to a queue along with 

the units (%T). The display has two icons (3) that are used to indicate 

the use of access code and offset mode. In addition, mode arrows 

(4) are used to indicate the current instrument operating mode; AUTO 

(normal operation), CAL (calibration) and CONFIG (configuration).

2.5 Touch Pad

Figure 2 illustrates the touch pad. 

MODE/EXIT

 , 

 and 

. The 

MODE/EXIT

 button is used to cycle between the three operational 

modes of the instrument: 

CAL

CONFIG

 and 

AUTO

 (Measurement) 

Mode. The 

  button enters the option (or mode that is highlighted or 

chosen). The 

 and 

 buttons are used to change settings.

Figure 1 – Display used in the instrument

All items used on the display are shown in this figure

3.0 Installation and Commissioning

Prior to use for the first time, the supplied desiccant pouch 

will need to be installed. Refer to section 

9.2 Replacing or 

Installing the Desiccant Pouch.

3.1 Mounting and Site Selection

The instrument is designed for wall mounting. For ease of service there 

should be about 20 cm (8") free area above the instrument; this will ensure 

enough room for calibration and cuvette maintenance. Choose a location 

that is easily accessible for operation and service and ensure that the 

front display rests at eye level. The overall mounting dimensions of the 

instrument are shown in Figure 3. The recommended mounting screws 

are M6 (¼") for the instrument enclosure and M4 (

3

16

") for the field terminal 

box. The AccUView LED is designed to have the field terminal box cradled 

under the sensor portion of the instrument. It is recommended that the 

field terminal box be mounted first, and then the rest of the instrument be 

mounted on top. The template included on the last page of this manual 

may be used to establish mounting-hole locations.

Figure 2: The AccUView LED touch pad

2.6 Vapor Purge

The AccUView LED is equipped with a continuous vapor purge system. 

A replaceable desiccant pouch in the lower portion of the instrument 

dries the air. System heat is used to warm the air. A fan inside the 

instrument continuously circulates heated dry air around the optical well 

and the flow through cuvette. This feature eliminates the need for a dry 

purge line.
The AccUView LED monitors the replaceable desiccant pouch 

condition continuously. The LCD display will show 

DESC

 on the 

lower line in the event that the desiccant pouch needs replacement. 

Replacement desiccant pouches are available from HF 

scientific, or the local representative (Part # 21555R). Refer 

to section 

9.2 Replacing or Installing the Desiccant Pouch

Figure 3: Overall Mounting Dimensions of the Instrument

It is critical that the instrument be mounted as close as possible to the 

sampling point to ensure a quick response time (within 2-3 meters (6-10 ft) 

of the sampling point).

Provide at least 200mm  

(8 inches) of free space 

for easy removal of the 

flow head and insertion of 

the calibration standards

All dimensions are in 

millimeters (inches)

These 4 holes will 

accept up to M6  

(1/4 inch) hardware

207.65

(8.175)

50.80

(2.000)

90.17

(3.500)

These 2 holes will 

accept up to M4  

(

3

16

 inch) hardware

87.33

(3.438)

30.17

(1.188)

147.66

(5.813)

131.16

(5.164)

347.16

(13.668)

61.77

(2.432)

196.85

(7.750)

Summary of Contents for HF scientific AccUView LED

Page 1: ...e Monitor IS HF AccUViewLED WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage o...

Page 2: ...d and passed EN 61326 1 2013 Manufacturer s Name HF scientific inc Manufacturer s Address 16260 Airport Park Drive Suite 140 Fort Myers Florida 33913 Importer s Name Importer s Address Type of Equipme...

Page 3: ...etting the Error Level 11 7 3 Configuring the RS 485 Port 11 7 3 1 Setting the Baud Rate 11 7 3 2 Setting the Address 11 7 4 Configuring the Alarms 11 7 4 1 Alarm 1 12 7 4 2 Alarm 2 12 7 5 Enabling th...

Page 4: ...indicates the items in the monitor shipment Item Quantity AccUView LED Monitor c w Field Terminal Box and Flow Through Assembly 1 Instruction Manual 1 Desiccant Pack 1 Quartz Cuvette with Ultrasonic...

Page 5: ...recommended mounting screws are M6 for the instrument enclosure and M4 3 16 for the field terminal box The AccUView LED is designed to have the field terminal box cradled under the sensor portion of...

Page 6: ...Model 28042 is equipped with a flow switch For this instrument the user connections are equipped with push to connect connections The suitable tubing is polymer or copper tubing with a inch OD These c...

Page 7: ...e While making connections refer to Figure 5 The AccUView LED is not supplied with a power cord 3 3 2 RS 485 The RS 485 Modbus or special protocol half duplex 2 wire digital interface operates with di...

Page 8: ...or the digital output depending on the options selected A flashing display may occur after warm up This is an indication that either the ambient or the water temperature or has changed rapidly During...

Page 9: ...e calibration 7 The upper display will count down the progress of the calibration step 8 The instrument will return to AUTO mode at the end the calibration 9 The shutoff clamp should now be released t...

Page 10: ...n of the AccUView LED will be resumed The AccUView LED will always reset to T and factory defaults after restoring NOTICE 7 0 Instrument Configuration CONFIG mode The instrument has been designed to p...

Page 11: ...rm 1 The alarm function HI LO OFF or Error 2 The alarm set point level at which the alarm activates 3 The delay time for the alarm the time that the set point must be exceeded prior to alarm activatio...

Page 12: ...tionalities is selected a prompt to set the delay times and the set point as with Alarm 1 will be displayed 7 5 Enabling the Security Access The instrument is equipped with a security access If this o...

Page 13: ...The factory 6 second update will extend the LED life to over 5 years The 60 second setting increases LED life to about 14 years but this will delay all updates Change the selected update period using...

Page 14: ...ust be used for the system to operate correctly The system works by sending an ultrasonic frequency through spring connections into a piezo transducer bonded to the bottom of a flow through cuvette Th...

Page 15: ...ed shutoff clamp unscrewing the old cuvette and replacing with a fresh clean one Recalibration will be required for accurate operation 9 2 Replacing or Installing the Desiccant Pouch The AccUView LED...

Page 16: ...the fault will be shown on the lower row of the display 10 2 System FAIL Message Normally this condition indicates that the instrument will require servicing Contact either the HF scientific Technica...

Page 17: ...ly 28474 AccUView LED Version with Flow Alarm Factory Installed Provides an alarm if sample flow is interrupted 28042 Tubing Kit Containing 1 shutoff clamp 1 backpressure valve 2 connecting tubings wi...

Page 18: ...3 500 87 33 3 438 30 17 1 188 147 66 5 813 Note 1 See the mounting instructions in the manual for mounting hardware sizes 2 Provide at least 200mm 8 inches of free space above the sensor for easy rem...

Page 19: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 20: ...ompany shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if t...

Reviews: