Watts HF scientific AccUView LED Instruction For Operation And Maintenance Download Page 9

9

5.0 Instrument Calibration

The instrument was calibrated and tested prior to leaving the factory. 

Therefore, it is possible to use the instrument directly out of the box. 

Under normal conditions, re-calibration is recommended at least once 

every month. To get the greatest accuracy, you may need to calibrate 

once per week.
Two cuvettes are supplied with the instrument so that they can be 

exchanged with each other. One cuvette should always be kept clean 

for replacement when calibrating.
The instrument is designed to store the last reading prior to calibrating. 

Relay contacts are held at the last valid condition and will not change 

state while the instrument is in the calibration mode. The 4-20 mA is 

also held at this stored reading and will not change until is calibration 

is complete. Thirty seconds after completion of the calibration the 

AccUView LED will start will start taking online measurements which will 

be averaged with the stored reading (depending on setting). In order 

to prevent disruptions in data the operator should release the clamp to 

allow the sample water to flush out the calibration water as soon as the 

calibration is completed.
While in the calibration mode, the instrument has a time-out feature 

that automatically returns the system operation to the 

AUTO

 mode 

after a fifteen (15) minute period of inactivity.

5.1 100 %T Calibration

1.  Select the calibration function of the instrument by pressing 

the 

MODE/EXIT

 button once. The arrow beside 

CAL

 will be 

illuminated on the display. The lower display shows alternating 

100

 (indicating that the 100%T standard is requested) and 

 . 

The last reading has now been stored and the equivalent reading is 

being output on the 4-20mA signal. This output will be held after the 

instrument returns to AUTO, allowing time for the flow to be restarted. 

2.  Close off the flow of water using the shutoff clamp.

3.  Remove the flow through cuvette, and retain for future cleaning  

and use.

4.  Fill the second clean cuvette with 100% T calibration fluid and 

reinstall the filled cuvette tightly on the flow through unit.

5.   

Wipe flow through unit dry

, install and lock in place.

6.  Press the 

  button to start the calibration.

7.  The upper display will count down the progress of the calibration step.
8.  The instrument will return to 

AUTO

 mode at the end the calibration.

9.  The shutoff clamp should now be released to restore sample flow.
10.  

Do not move flow head after calibrating.

 

Before starting the calibration procedure, be sure to turn off  

the instrument’s ultrasonic cleaning option. See section 7.10  

for instructions.

NOTICE

5.2 ABS Calibration

1.  Select the calibration function of the instrument by pressing 

the 

MODE/EXIT

 button once. The arrow beside 

CAL

 will be 

illuminated on the display. The lower display shows alternating 

0

 

(indicating that the 0.000ABS standard is requested) and 

 . Once 

in the calibration menu the last reading has now been stored and 

the equivalent reading is being output on the 4-20mA signal. This 

output will be held after the instrument returns to AUTO, allowing 

time for the flow to be restarted.

2.  Close off the flow of water using the shutoff clamp.

3.  Remove the flow through cuvette, and retain for future cleaning and use.
4.  Fill the second clean cuvette with 100% T/0.000 ABS calibration fluid 

and reinstall the filled cuvette on the flow through unit.

5.   

Wipe flow through unit dry

, install and lock in place.

6.  Press the 

  button to start the calibration.

7.  The upper display will count down the progress of the calibration step.
8.  The instrument will return to 

AUTO

 mode at the end the calibration.

9.  The shutoff clamp should now be released to restore sample flow.
10.  

Do not move flow head after calibrating.

Since the instrument stored the reading previous to calibration, the 

previous reading may appear after calibration for at least 30 seconds. The 

reading will then update to the current reading. How fast this upgrades 

occurs depends on the 

CONFIG

 settings for 

RESP

 (response time) and 

UV-T

 (lamp duty cycle).

There is no need to recalibrate when changing ranges from %T to ABS or 

the reverse.

It is recommended that the flow through unit be removed for as 

short an interval as possible to prevent premature saturation of 

the desiccant.

NOTICE

CAUTION

!

The UV lamp is of very low power but is recommend to avoid 

looking directly into the optical well while the flow through unit 

is removed.

Before starting the calibration procedure, be sure to turn off  

the instrument’s ultrasonic cleaning option. See section 7.10  

for instructions.

NOTICE

Summary of Contents for HF scientific AccUView LED

Page 1: ...e Monitor IS HF AccUViewLED WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage o...

Page 2: ...d and passed EN 61326 1 2013 Manufacturer s Name HF scientific inc Manufacturer s Address 16260 Airport Park Drive Suite 140 Fort Myers Florida 33913 Importer s Name Importer s Address Type of Equipme...

Page 3: ...etting the Error Level 11 7 3 Configuring the RS 485 Port 11 7 3 1 Setting the Baud Rate 11 7 3 2 Setting the Address 11 7 4 Configuring the Alarms 11 7 4 1 Alarm 1 12 7 4 2 Alarm 2 12 7 5 Enabling th...

Page 4: ...indicates the items in the monitor shipment Item Quantity AccUView LED Monitor c w Field Terminal Box and Flow Through Assembly 1 Instruction Manual 1 Desiccant Pack 1 Quartz Cuvette with Ultrasonic...

Page 5: ...recommended mounting screws are M6 for the instrument enclosure and M4 3 16 for the field terminal box The AccUView LED is designed to have the field terminal box cradled under the sensor portion of...

Page 6: ...Model 28042 is equipped with a flow switch For this instrument the user connections are equipped with push to connect connections The suitable tubing is polymer or copper tubing with a inch OD These c...

Page 7: ...e While making connections refer to Figure 5 The AccUView LED is not supplied with a power cord 3 3 2 RS 485 The RS 485 Modbus or special protocol half duplex 2 wire digital interface operates with di...

Page 8: ...or the digital output depending on the options selected A flashing display may occur after warm up This is an indication that either the ambient or the water temperature or has changed rapidly During...

Page 9: ...e calibration 7 The upper display will count down the progress of the calibration step 8 The instrument will return to AUTO mode at the end the calibration 9 The shutoff clamp should now be released t...

Page 10: ...n of the AccUView LED will be resumed The AccUView LED will always reset to T and factory defaults after restoring NOTICE 7 0 Instrument Configuration CONFIG mode The instrument has been designed to p...

Page 11: ...rm 1 The alarm function HI LO OFF or Error 2 The alarm set point level at which the alarm activates 3 The delay time for the alarm the time that the set point must be exceeded prior to alarm activatio...

Page 12: ...tionalities is selected a prompt to set the delay times and the set point as with Alarm 1 will be displayed 7 5 Enabling the Security Access The instrument is equipped with a security access If this o...

Page 13: ...The factory 6 second update will extend the LED life to over 5 years The 60 second setting increases LED life to about 14 years but this will delay all updates Change the selected update period using...

Page 14: ...ust be used for the system to operate correctly The system works by sending an ultrasonic frequency through spring connections into a piezo transducer bonded to the bottom of a flow through cuvette Th...

Page 15: ...ed shutoff clamp unscrewing the old cuvette and replacing with a fresh clean one Recalibration will be required for accurate operation 9 2 Replacing or Installing the Desiccant Pouch The AccUView LED...

Page 16: ...the fault will be shown on the lower row of the display 10 2 System FAIL Message Normally this condition indicates that the instrument will require servicing Contact either the HF scientific Technica...

Page 17: ...ly 28474 AccUView LED Version with Flow Alarm Factory Installed Provides an alarm if sample flow is interrupted 28042 Tubing Kit Containing 1 shutoff clamp 1 backpressure valve 2 connecting tubings wi...

Page 18: ...3 500 87 33 3 438 30 17 1 188 147 66 5 813 Note 1 See the mounting instructions in the manual for mounting hardware sizes 2 Provide at least 200mm 8 inches of free space above the sensor for easy rem...

Page 19: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 20: ...ompany shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if t...

Reviews: