2.0 Introduction to the Unit

2.1 Overview

The AccUView LED process transmittance monitor allows for the UV 

transmittance or absorbance measurement of process water on-line. 

A unique feature of the instrument is the ultrasonic cleaning. 

Refer to section 8.2 for more information.

2.2 Specifications

Measurement Range

0 – 100.0 %T 

0 – 2.0 ABS 

Repeatability

± 1.0 %T

 ± 0.002 ABS

Resolution

0.1 % T

0.0004 ABS

Accuracy

± 2.0 % T 

± 0.008 ABS 

Response Time

Adjustable Averaging 

Path Length

22 mm (Readings corrected to 10 mm) 

Display

Multi-Line Liquid Crystal Backlit Display 

Alarms

Two Programmable, 120-240VAC 2A Form C Relay

Analog Output 

Powered loop isolator 4-20 mA, 600 Ω drive 

Communications Port 

Optional bi-directional RS-485 with Modbus 

Maximum Water Pressure 

1380 kPa (200 psi.) with integral pressure regulator 

Flow Rate 

100 ml/min. – 1.5 liter/min. (.026-.40 Gal/min) 

Operating Temperature 

0°C – 50°C (32°F – 122°F) 

Wetted Materials 

Nylon, Quartz, Silicone, Polypropylene, Stainless Steel 

Sample Temperature Range 

0°C – 50°C (34°F – 122°F) 

Power Supply 

100– 240 VAC, 47 – 63 Hz, 80VA 

Insulation Rating 

Double Insulated, Pollution Degree 2, Overvoltage Category II 

Environmental Conditions

Not recommended for outdoor use. 

Altitude up to 2000 meters 

Up to 95 % RH (non-condensing) 

Enclosure

ABS

Regulatory Compliance 

and Certifications

CE Approved, Listed to UL 61010-1 3rd edition: 2012 

Certified to CAN/CSA C22.2 No. 61010-1: 2012 

EMC: EN61326-1: 2013 

Shipping Weight

2.5 kg (5.5 lbs.) 

Warranty

1 Year from date of shipment

2.3  Unpacking and Inspection of the 

Instrument and Accessories

The table below indicates the items in the monitor shipment.

Item

Quantity

AccUView LED Monitor c/w Field Terminal Box and Flow Through Assembly

1

Instruction Manual

1

Desiccant Pack

1

Quartz Cuvette with Ultrasonic Transducer (Single Pack)

1

100 % T Calibration Standard

1

Tubing Kit *

1

* Note - Tubing Kit includes:

1-shutoff clamp

1-backpressure valve

2-connecting tubings with fittings for flow through assembly

1-drain vent screw (used in pressurized systems)

Refer to Figure 4.

Remove the instrument from the packing carton. Carefully inspect all 

items to ensure that no visible damage has occurred during shipment. 

If the items received do not match the order, please immediately contact 

the local distributor or the HF scientific Customer Service department.

This symbol identifies hazards which, if not avoided, 

could result in minor or moderate injury or damage 

to the equipment. 

This symbol identifies important information, 

practices or actions.

This pictorial alert you to the need read the manual, 

possibly at a different section.

This pictorial alerts you to electricity, electrocution 

and shock hazards. 

1.0  Understanding Safety  

Information

This manual contains basic instructions that must be followed during 

the commissioning, operation, care and maintenance of the instrument. 

The safety protection provided by this equipment may be impaired if it 

is commissioned and/or used in a manner not described in this manual. 

Consequently, all responsible personnel must read this manual prior to 

working with this instrument.
In certain instances icons or symbols have been added to give further 

clarification to the instructions. Refer to the 

Table of Contents

 to easily 

find specific topics and to learn about unfamiliar terms.

1.1 Symbols Used in this manual

CAUTION

!

NOTICE

Summary of Contents for HF scientific AccUView LED

Page 1: ...e Monitor IS HF AccUViewLED WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage o...

Page 2: ...d and passed EN 61326 1 2013 Manufacturer s Name HF scientific inc Manufacturer s Address 16260 Airport Park Drive Suite 140 Fort Myers Florida 33913 Importer s Name Importer s Address Type of Equipme...

Page 3: ...etting the Error Level 11 7 3 Configuring the RS 485 Port 11 7 3 1 Setting the Baud Rate 11 7 3 2 Setting the Address 11 7 4 Configuring the Alarms 11 7 4 1 Alarm 1 12 7 4 2 Alarm 2 12 7 5 Enabling th...

Page 4: ...indicates the items in the monitor shipment Item Quantity AccUView LED Monitor c w Field Terminal Box and Flow Through Assembly 1 Instruction Manual 1 Desiccant Pack 1 Quartz Cuvette with Ultrasonic...

Page 5: ...recommended mounting screws are M6 for the instrument enclosure and M4 3 16 for the field terminal box The AccUView LED is designed to have the field terminal box cradled under the sensor portion of...

Page 6: ...Model 28042 is equipped with a flow switch For this instrument the user connections are equipped with push to connect connections The suitable tubing is polymer or copper tubing with a inch OD These c...

Page 7: ...e While making connections refer to Figure 5 The AccUView LED is not supplied with a power cord 3 3 2 RS 485 The RS 485 Modbus or special protocol half duplex 2 wire digital interface operates with di...

Page 8: ...or the digital output depending on the options selected A flashing display may occur after warm up This is an indication that either the ambient or the water temperature or has changed rapidly During...

Page 9: ...e calibration 7 The upper display will count down the progress of the calibration step 8 The instrument will return to AUTO mode at the end the calibration 9 The shutoff clamp should now be released t...

Page 10: ...n of the AccUView LED will be resumed The AccUView LED will always reset to T and factory defaults after restoring NOTICE 7 0 Instrument Configuration CONFIG mode The instrument has been designed to p...

Page 11: ...rm 1 The alarm function HI LO OFF or Error 2 The alarm set point level at which the alarm activates 3 The delay time for the alarm the time that the set point must be exceeded prior to alarm activatio...

Page 12: ...tionalities is selected a prompt to set the delay times and the set point as with Alarm 1 will be displayed 7 5 Enabling the Security Access The instrument is equipped with a security access If this o...

Page 13: ...The factory 6 second update will extend the LED life to over 5 years The 60 second setting increases LED life to about 14 years but this will delay all updates Change the selected update period using...

Page 14: ...ust be used for the system to operate correctly The system works by sending an ultrasonic frequency through spring connections into a piezo transducer bonded to the bottom of a flow through cuvette Th...

Page 15: ...ed shutoff clamp unscrewing the old cuvette and replacing with a fresh clean one Recalibration will be required for accurate operation 9 2 Replacing or Installing the Desiccant Pouch The AccUView LED...

Page 16: ...the fault will be shown on the lower row of the display 10 2 System FAIL Message Normally this condition indicates that the instrument will require servicing Contact either the HF scientific Technica...

Page 17: ...ly 28474 AccUView LED Version with Flow Alarm Factory Installed Provides an alarm if sample flow is interrupted 28042 Tubing Kit Containing 1 shutoff clamp 1 backpressure valve 2 connecting tubings wi...

Page 18: ...3 500 87 33 3 438 30 17 1 188 147 66 5 813 Note 1 See the mounting instructions in the manual for mounting hardware sizes 2 Provide at least 200mm 8 inches of free space above the sensor for easy rem...

Page 19: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 20: ...ompany shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if t...

Reviews: