Post-Injection Volume Kit
instructions
Warning:
To avoid personal contamination with biologically hazardous or toxic compounds,
wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Requirement:
Wear clean, chemical-resistant, powder-free gloves when performing
this procedure.
Required tools and materials
• Chemical-resistant, powder-free gloves
• 1/4-inch open-end wrench
• Post-injection loop kit
To install or replace a post-injection loop:
1.
Power-off the sample manager.
2.
Open the fluidics compartment door.
3.
Fully insert the post-injection loop tubing into port 6 of the injection valve, and then thread
the fitting into the port.
4.
Holding the extension post-injection loop tubing against the bottom of the port, finger-
tighten the post-injection loop fitting.
5.
Use the 1/4-inch open-end wrench to tighten the post-injection loop fitting an additional 1/6-
turn for existing fittings, or 3/4-turn for a new fitting.
Figure A–1: Post-injection loop
1
2
3
April 4, 2018, 715005702 Rev. A