Technical data
Lyla | Instructions for Use | RIN-210 ENG Rev02 | 15/09/2022 | © W&H Sterilization Srl
101
MAXIMUM LOAD FOR INSTRUMENTS
Note
: the load given includes the trays, the containers and everything is put into the chamber, with the sole exclusion of the tray rack.
Instruments
Bagged
Unwrapped
Porous
Lyla 17
Lyla 22
Lyla 17
Lyla 22
Lyla 17
Lyla 22
S Unwrapped 134
-
-
4.0 kg
4.0 kg
-
-
S Wrapped 134
4.0 kg
4.0 kg
4.0 kg
4.0 kg
0.5 kg
0.5 kg
S Gentle 121
4.0 kg
4.0 kg
4.0 kg
4.0 kg
0.5 kg
0.5 kg
S Prion 134
4.0 kg
4.0 kg
4.0 kg
4.0 kg
0.5 kg
0.5 kg
B Universal 134 (only with AC)
3.5/5.5 kg
1
4.0/6.0 kg
2
3.5/5.5 kg
1
4.0/6.0 kg
2
0.5/1.5 kg
3
0.5/2.0 kg
4
B Universal 134 ECO mode (only with AC)
0.5 kg
0.5 kg
0.5 kg
0.5 kg
0.5 kg
0.5 kg
B Prion / Extended 134 (only with AC)
3.5/5.5 kg
1
4.0/6.0 kg
2
3.5/5.5 kg
1
4.0/6.0 kg
2
0.5/1.5 kg
3
0.5/2.0 kg
4
B Extended 134 ECO mode (only with AC)
0.5 kg
0.5 kg
0.5 kg
0.5 kg
0.5 kg
0.5 kg
B Universal 121 (only with AC)
3.5/5.5 kg
1
4.0/6.0 kg
2
3.5/5.5 kg
1
4.0/6.0 kg
2
0.5/1.5 kg
3
0.5/2.0 kg
4
1
: duration reported in the table refers to 3.5 kg. It is possible to sterilize up to 5.5 kg by drying time extension.
2
: duration reported in the table refers to 4.0 kg. It is possible to sterilize up to 6.0 kg by drying time extension.
3
: duration reported in the table refers to 0.5 kg. It is possible to sterilize up to 1.5 kg by drying time extension.
4
: duration reported in the table refers to 0.5 kg. It is possible to sterilize up to 2.0 kg by drying time extension.
WARNING! Never exceed the maximum load weight limits as specified in the table as this could impair the sterilization process.