Maintenance
64
Lyla | Instructions for Use | RIN-210 ENG Rev02 | 15/09/2022 | © W&H Sterilization Srl
CLEANING THE CHAMBER AND THE CHAMBER ACCESSORIES
Proceed as follows:
1
Remove the trays, the chamber rack and the steam diffuser plate.
2
Clean the chamber with a damp sponge and a mild detergent solution paying attention not to bend or damage the temperature probe
inside the sterilizer chamber.
3
Rinse with water.
4
Clean the trays, the chamber rack and the steam diffuser plate with a damp sponge and a mild detergent solution.
5
Rinse with water.
6
Reposition all pieces of the chamber accessories properly.
WARNING! Ensure that the steam diffuser plate is correctly placed and engaged, as this is essential for the sterilization process.
Note
: the trays, the tray holder and the steam diffuser plate may also be cleaned in a washer disinfector.