Diagnostics
88
Lyla | Instructions for Use | RIN-210 ENG Rev02 | 15/09/2022 | © W&H Sterilization Srl
RESET THE SAFETY THERMOSTAT
The sterilizer is fitted with a safety thermostat to prevent it from overheating. If the safety thermostat operates because of too high temperatures,
the error 240 or a timeout error is displayed. The thermostat must be reset manually. Proceed as follows:
1
Switch the sterilizer OFF and remove the mains cable.
2
Wait for the sterilizer to cool down.
3
Open the chamber door.
4
Remove the dust filter and move the sterilizer closer to the shelf edge.
5
Push the reset button of the thermostat switch: a click sound indicates that the
thermostat switch has been reset.