Maintenance
Lyla | Instructions for Use | RIN-210 ENG Rev02 | 15/09/2022 | © W&H Sterilization Srl
65
CLEANING THE CHAMBER FILTER
Proceed as follows:
1
Allow the sterilization chamber to cool down.
2
Empty the sterilizer chamber by removing the trays and the rack.
3
Turn the filter cap at the back of the chamber (bottom/center) counter-clockwise.