Diagnostics
84
Lyla | Instructions for Use | RIN-210 ENG Rev02 | 15/09/2022 | © W&H Sterilization Srl
Code
Description
Actions
3xx
Check the door gasket. Clean or replace it if necessary.
Repeat the cycle.
If the problem persists, call service.
Clean the chamber face side.
Clean the chamber filter.
Check if chamber filter is properly locked in the cap.
Check the load does not exceed the MAXIMUM WEIGHT LIMITS.
4xx
Clean water error (bad quality or low water level). Drain and refill the clean water tank.
Repeat the cycle.
If the problem persists, call service.
5xx
Check if there are hurdles on the door locking area (chamber rack, loads, objects, …).
Repeat the cycle.
If the problem persists, call service.
Check the door gasket (wrong placed).
Check if the door can move freely without touching the trays or the load when closing.
Switch the sterilizer OFF and ON.
990
The cycle has been aborted by the user.
Re-process the load.