22
Lyla | Instructions for Use | RIN-210 ENG Rev02 | 15/09/2022 | © W&H Sterilization Srl
Getting started
The installation of the sterilizer shall be performed by two authorized
technicians using PPE (Personal Protective Equipment) according to
applicable standards.
The electrical power supply of the sterilizer must fulfill all applicable
standards in the country of use, and must comply with the data label
on the back of the sterilizer.
WATER CONNECTIONS
The sterilizer clean water tank can be filled manually by the user or
automatically with a water supply system (optional). The water
supply system must deliver demineralized or distilled water meeting
the specifications listed in these instructions. Do not add any
chemical/additive to the water.
The manufacturer’s warranty is void if the sterilizer was used with
water containing either chemical additives, or contaminant levels
exceeding those listed in these instructions. See "Feed water
specifications (EN 13060)" on page 109.
Notice
: the maintenance of the external water filling system must be
done in exact accordance with the Instructions for Use given with the
relevant system.
WI-FI CONNECTION (OPTIONAL)
For the Wi-Fi connection proceed as follows:
1
Insert the Wi-Fi dongle key in the USB port.
2
Read the Instructions for Use provided with the Wi-Fi dongle key.
INSTALLING THE STERILIZER
WARNING! In case of sterilizer malfunctions immediately
unplug the sterilizer and call for service. Do not attempt to
repair the sterilizer by yourself.
Notice
:
Please ensure that all installation requirements are met before
plugging the sterilizer. See "Connection diagrams" on page 108.
No other devices should be connected to the sterilizer power panel
circuit.
1
Place the sterilizer on a sturdy, flat and level surface.
2
Open the chamber door, remove all items from the sterilizer
chamber except the chamber rack. Remove all plastic covers
from trays.
3
Connect the auto-fill and auto-drain tubes in the rear of the
sterilizer.
4
Connect the Ethernet cable or the Wi-Fi dongle key in the rear of
the sterilizer.
5
Attach the power cord to the socket in the rear of the sterilizer
and route the cord through the cable guides.
6
Connect the power cord to a wall outlet. For power supply
requirements, see "Technical data" on page 105.
7
Ensure that the steam diffuser plate is correctly placed and
engaged, as this is essential for the sterilization process.