Maintenance
78
Lyla | Instructions for Use | RIN-210 ENG Rev02 | 15/09/2022 | © W&H Sterilization Srl
5
Insert the new seal and press first up and down.
6
Press left and right and then the entire seal circumference to ensure its perfect
placement.
Notice
: a steam discharge can damage the sterilizer plastic parts. Ensure the seal
does not stick out.
7
Wipe any residual water and run a Vacuum and Helix test to check for perfect
tightness of the seal. See "Vacuum test" on page 45 and "Helix test" on page 43.