Technical data
Lyla | Instructions for Use | RIN-210 ENG Rev02 | 15/09/2022 | © W&H Sterilization Srl
97
STANDARD STERILIZATION CYCLES AVAILABLE
There are seven sterilization cycles available, all of them complying with the European Standard EN13060:
n
four S-type cycles
n
three B-type cycles (optional, activated with B cycles activation code. See "B-type cycles activation" on page 51)
Cycle type
Cycle name
Purpose
S
S Unwrapped 134
Very fast processing of unwrapped instruments.
S Wrapped 134
Designed for wrapped instruments and porous load.
S Gentle 121
Designed for heat-sensitive items and porous load wrapped or unwrapped, such as textiles and plastics.
S Prion 134
Features an extended sterilization time, if this is required for your load or mandated in your country.
B
(available only with Activation Code)
B Universal 134
For all your general items like hand instruments, handpieces, forceps, etc.
B Prion / Extended 134 *
Features an extended sterilization time, if this is required for your load or mandated in your country.
B Universal 121
For all items that cannot withstand the high temperatures of the 134 cycles, such as textiles and plastics.
*
: cycle name might be different, depending on country requirements.
STERILIZATION CYCLE DATA (S-TYPE CYCLES)
Sterilization cycles (S-type cycles)
S Unwrapped 134
S Wrapped 134
S Gentle 121
S Prion 134
Sterilization temperature
134 °C (273 °F)
134 °C (273 °F)
121 °C (250 °F)
134 °C (273 °F )
Sterilization pressure
3.03 bar
2.03 bar (g)
3.03 bar
2.03 bar (g)
2.04 bar
1.04 bar (g)
3.03 bar
2.03 bar (g)
Duration of the Plateau/Sterilization phase
4'
4’
20'30"
18'30"