Maintenance
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
65
REPLACING THE HEPA FILTER
Notice
: the HEPA filter needs to be replaced every 1200 sterilization cycles or once a year, whichever come first. A replacement message alerts
when replacement is due. If the consumables is replaced prior to the message, you have to reset the consumable cycle counter.
Proceed as follows:
1
Open the sterilizer door.
2
Unscrew the HEPA filter by hand (counter-clockwise).
3
Screw on the new HEPA filter (clockwise) and tighten it snug.
Note
: tap
to see the animated replacement procedure.