CUSTOM STERILIZATION PROGRAMS
WARNING! The custom programs are not FDA cleared and it is the responsibility of the user to validate these programs.
The sterilizer provides the following custom programs. These programs are disabled by default.
Program
Sterilization Temperature
Sterilization Time
(settable)
Drying time
(settable)
Custom A
1
250 °F (121 °C)
15–60 minutes
30–60 minutes
Custom B
1
270 °F (132 °C )
4–30 minutes
3–60 minutes
Custom C
1
273 °F (134 °C)
3–30 minutes
3–60 minutes
1
: these are profiles defined by ANSI/AAMI ST55 as dynamic-air-removal (steam-flush pressure-pulse). These profiles can be enabled by the user. Some of the
parameters can be adjusted by the user according to his needs (Plateau/Sterilization phase, dry time, etc.). These cycles are not FDA cleared.
STERILIZATION PROGRAM OPTIONS (200–240 V AC VERSION)
WARNING! The sterility of items processed unwrapped is compromised on exposure to non-sterile environments. Ensure that items
are dry when removed from the sterilizer. If the default drying time is not adequate for the load to be sterilized, additional drying
time can be added (see "Sterilization cycle management" on page 48). To load properly the sterilizer see "Prepare the sterilizer" on
page 47.